15-4
NOTE
To extend the lifetime of the watertrap and module, disconnect the watertrap and set the operating mode
to standby mode when CO
2
monitoring is not required.
15.3.2
Using a Microstream CO
2
Module
Connect the sampling line to the module and then connect the CO
2
components as shown below. The message [
CO2
Sensor Warmup
] is displayed. After warm-up, you can perform CO
2
measurements. The message [
CO2 Sensor Warmup
]
is displayed.
15.3.3
Using a Mainstream CO
2
Module
1.
Connect the sensor to the module. The message [
CO2 Sensor Warmup
] is displayed.
2.
After warm-up is finished, connect the transducer to the airway adapter.
3.
Perform a zero calibration per
15.9 Zeroing the Sensor
.
4.
After the zero calibration is finished, connect the airway as shown below.
5.
Make sure there are no leakages in the airway and then start a measurement.
Sampling line
Connector for
sampling line
Airway adapter
Sensor
Connect to patient
Connector to monitor
Summary of Contents for BeneView T1
Page 1: ...BeneView T1 Patient Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 20: ...12 FOR YOUR NOTES...
Page 38: ...2 14 FOR YOUR NOTES...
Page 56: ...4 8 FOR YOUR NOTES...
Page 64: ...5 8 FOR YOUR NOTES...
Page 116: ...8 30 FOR YOUR NOTES...
Page 130: ...11 8 FOR YOUR NOTES...
Page 138: ...12 8 FOR YOUR NOTES...
Page 140: ...13 2 FOR YOUR NOTES...
Page 174: ...16 12 FOR YOUR NOTES...
Page 190: ...18 10 FOR YOUR NOTES...
Page 206: ...21 6 FOR YOUR NOTES...
Page 246: ...A 20 FOR YOUR NOTES...
Page 274: ...D 8 FOR YOUR NOTES...
Page 285: ...G D Declarat tion of C Conform G 1 mity...
Page 286: ...P N 046 005297 00 15 0...