A-17
Rise time (module PN: 115-027545-00)
Measured with a DRYLINE II neonatal watertrap and a 2.5-meter neonatal sampling
line:
<330 ms@90 ml/min.
Measured with a DRYLINE II adult watertrap and a 2.5-meter adult sampling line:
<300 ms@120 ml/min
awRR measurement range
0 to 120 rpm
awRR measurement precision
±2 rpm
Apnea time
10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s
Note: The response time is the sum of the rise time and the delay time.
Effect of interference gases on CO
2
measurements
Gas
Concentration (%)
Quantitive effect*
N
2
O
≤60
±1 mmHg
Hal
≤4
Sev
≤5
Iso
≤5
Enf
≤5
Des
≤15
±2 mmHg
*: means an extra error should be added in case of gas interference when CO
2
measurements are performed between 0 -
40mmHg.
Inaccuracy specifications are affected by the breath rate and I:E change. The end-tidal gas reading is within specification for
breath rate below 15BPM and I:E ratio smaller than 1:1 relative to the gas readings without breath.
Alarm limit
Range
Step
EtCO
2
High (mmHg)
(low limit + 2) to 99
1
EtCO
2
Low (mmHg)
1 to (high limit - 2)
FiCO
2
High (mmHg)
1 to 99
awRR High (rpm)
Adult, pediatric:
Neonate:
(low limit + 2) to 100
(low limit + 2) to 150
1
awRR Low (rpm)
0 to (high limit - 2)
Microstream CO
2
Module
Standard
Meet standard of ISO 21647
CO
2
Measurement range
0 to 99 mmHg
Accuracy*
0 to 38 mmHg:
39 to 99 mmHg:
±2 mmHg
±5% of the 0.08% of (the reading-38)
Accuracy drift
Meet the requirement for measurement accuracy within 6 hours
* Accuracy applies for respiration rate up to 80 rpm. For respiration rate above 80 rpm, the accuracy is 4 mmHg or ±12% of
the reading, whichever is greater, for EtCO
2
exceeding 18 mmHg. For respiration rate above 60 rpm, the above accuracy can
be achieved by using the CapnoLine H Set for Infant/Neonatal. In the presence of interfering gases, the above accuracy is
maintained to within 4%.
Resolution
1 mmHg
Sample flow rate
7.5
15
50
−
+
ml/min
Initialization time
30 s (typical)
Summary of Contents for BeneView T1
Page 1: ...BeneView T1 Patient Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 20: ...12 FOR YOUR NOTES...
Page 38: ...2 14 FOR YOUR NOTES...
Page 56: ...4 8 FOR YOUR NOTES...
Page 64: ...5 8 FOR YOUR NOTES...
Page 116: ...8 30 FOR YOUR NOTES...
Page 130: ...11 8 FOR YOUR NOTES...
Page 138: ...12 8 FOR YOUR NOTES...
Page 140: ...13 2 FOR YOUR NOTES...
Page 174: ...16 12 FOR YOUR NOTES...
Page 190: ...18 10 FOR YOUR NOTES...
Page 206: ...21 6 FOR YOUR NOTES...
Page 246: ...A 20 FOR YOUR NOTES...
Page 274: ...D 8 FOR YOUR NOTES...
Page 285: ...G D Declarat tion of C Conform G 1 mity...
Page 286: ...P N 046 005297 00 15 0...