7-6
7.4.5
Setting the Reminder Tones
When the alarm volume is set to zero, or the alarm is reset or switched off, the patient monitor issues a periodical
reminder tone.
1. Select
[
Main Menu
]
→
[
Maintenance >>
]
→
[
User Maintenance >>
]
→
enter the required password.
2. Select
[
Alarm Setup >>
] to enter the [
Alarm Setup
] menu.
3.
Set the [
Reminder Tones
] to [
On
], [
Off
] or [
Re-alarm
]. When [
Re-alarm
] is selected, the acknowledged
physiological alarms and technical alarms marked with “
√
” will be re-generated after the [
Reminder Interval
] if
the alarm condition persists.
To set the interval between reminder tones, select [
Reminder Interval
] and toggle between [
1min
], [
2min
] and [
3min
].
In addition, you can set the volume of alarm reminder tones. To set the volume of alarm reminder tones, select [
Main
Menu
]
→
[
Alarm Setup >>
]
→
[
Others
]. Then, select [
Reminder Vol
] and toggle between [
High
], [
Medium
] and [
Low
].
7.5
Understanding the Alarm Setup Menu
Select [
Main Menu
]
→
[
Alarm Setup >>
] to enter the [
Alarm Setup
], where you can:
Set alarm properties for all parameters.
Change ST alarm settings.
Change arrhythmia alarm settings.
Set the threshold for some arrhythmia alarms.
Change other settings.
Please refer to the
8 Monitoring ECG
for how to change ST alarm settings, how to change arrhythmia alarm settings and
how to set the threshold for some arrhythmia alarms.
Summary of Contents for BeneView T1
Page 1: ...BeneView T1 Patient Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 20: ...12 FOR YOUR NOTES...
Page 38: ...2 14 FOR YOUR NOTES...
Page 56: ...4 8 FOR YOUR NOTES...
Page 64: ...5 8 FOR YOUR NOTES...
Page 116: ...8 30 FOR YOUR NOTES...
Page 130: ...11 8 FOR YOUR NOTES...
Page 138: ...12 8 FOR YOUR NOTES...
Page 140: ...13 2 FOR YOUR NOTES...
Page 174: ...16 12 FOR YOUR NOTES...
Page 190: ...18 10 FOR YOUR NOTES...
Page 206: ...21 6 FOR YOUR NOTES...
Page 246: ...A 20 FOR YOUR NOTES...
Page 274: ...D 8 FOR YOUR NOTES...
Page 285: ...G D Declarat tion of C Conform G 1 mity...
Page 286: ...P N 046 005297 00 15 0...