C-2
ST Analysis
Item Name
Configurable
Default
In Config Mode
In Monitor Mode
ST Analysis
Yes
Yes
General, OR, ICU, NICU: Off
CCU: On
Alarm Yes
Yes Off
Alm Lev
Yes
Yes
Med
ST-X High
Yes
Yes
when ST Unit is mV:
when ST Unit is mm:
0.20
2.0
ST-X Low
Yes
Yes
when ST Unit is mV:
when ST Unit is mm:
-0.20
-2.0
ISO Yes
Yes
-80
ms
J Yes
Yes
48
ms
ST
Yes
Yes
J + 60 ms
X represents I, II, III, aVR, aVL, aVF, or V1 to V6.
QT/QTc Analysis
Item Name
Configurable
Default
In Config Mode
In Monitor Mode
QT Analysis
Yes
Yes
Off
QTc Formula
Yes
Yes
Hodges
Analysis Lead
Yes
Yes
All
Arrh. Analysis
Arrhythmia Threshold Settings (Mindray)
Item Name
Configurable
Default
In Config Mode
In Monitor Mode
PVCs High
Yes
Yes
10
Asys. Delay
Yes
Yes
5
Vtac Rate
Yes
Yes
130
Vtac PVCs
Yes
Yes
6
Multif. PVC’s Window
Yes
Yes
15
Tachy Yes
Yes
Adu:
Ped:
120
160
Brady Yes
Yes
Adu:
Ped:
50
75
Extreme Tachy
Yes
Yes
Adu:
Ped:
160
180
Extreme Brady
Yes
Yes
Adu:
Ped:
35
50
Vbrd Rate
Yes
Yes
40
Vbrd PVCs
Yes
Yes
5
Pause Time
Yes
Yes
2
Summary of Contents for BeneView T1
Page 1: ...BeneView T1 Patient Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 20: ...12 FOR YOUR NOTES...
Page 38: ...2 14 FOR YOUR NOTES...
Page 56: ...4 8 FOR YOUR NOTES...
Page 64: ...5 8 FOR YOUR NOTES...
Page 116: ...8 30 FOR YOUR NOTES...
Page 130: ...11 8 FOR YOUR NOTES...
Page 138: ...12 8 FOR YOUR NOTES...
Page 140: ...13 2 FOR YOUR NOTES...
Page 174: ...16 12 FOR YOUR NOTES...
Page 190: ...18 10 FOR YOUR NOTES...
Page 206: ...21 6 FOR YOUR NOTES...
Page 246: ...A 20 FOR YOUR NOTES...
Page 274: ...D 8 FOR YOUR NOTES...
Page 285: ...G D Declarat tion of C Conform G 1 mity...
Page 286: ...P N 046 005297 00 15 0...