C-5
C.1.3
PR
Item Name
Configurable
Default
In Config Mode
In Monitor Mode
Alarm Yes
Yes
On
Alm Lev
Yes
Yes
Med
HR/PR High
Yes
Yes
Adu: 120
Ped: 160
Neo: 200
HR/PR Low
Yes
Yes
Adul: 50
Ped: 75
Neo: 100
PR Source
Yes
Yes
Auto
Alm Source
Yes
Yes
Auto
Beat Vol
Yes
Yes
General, OR: 2
ICU, NICU, CCU: 1
C.1.4
SpO
2
Item Name
Configurable
Default
In Config Mode
In Monitor Mode
Alarm Yes
Yes
On
Alm Lev
Yes
Yes
SpO
2
Desat
Med
High
SpO
2
High
Yes
Yes
Adu, Ped:
Neo:
100
95
SpO
2
Low
Yes
Yes
90
Desat Yes
Yes
80
Sensitivity Yes Yes Mindray:
Masimo:
Med
Normal
Averaging (Masimo)
Yes
Yes
8 s
Sat-Seconds (Nellcor)
Yes
Yes
0 s
Sweep Yes
Yes
25
mm/s
NIBP Simul
Yes
Yes
Off
PI Zoom
Yes
Yes
No
C.1.5
Temp
Item Name
Configurable
Default
In Config Mode
In Monitor Mode
Alarm Yes
Yes
On
Alm Lev
Yes
Yes
Med
T1/T2 High (ºC)
Yes
Yes
38.0
T1/T2 Low (ºC)
Yes
Yes
35.0
TD High (ºC)
Yes
Yes
2.0
Summary of Contents for BeneView T1
Page 1: ...BeneView T1 Patient Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 20: ...12 FOR YOUR NOTES...
Page 38: ...2 14 FOR YOUR NOTES...
Page 56: ...4 8 FOR YOUR NOTES...
Page 64: ...5 8 FOR YOUR NOTES...
Page 116: ...8 30 FOR YOUR NOTES...
Page 130: ...11 8 FOR YOUR NOTES...
Page 138: ...12 8 FOR YOUR NOTES...
Page 140: ...13 2 FOR YOUR NOTES...
Page 174: ...16 12 FOR YOUR NOTES...
Page 190: ...18 10 FOR YOUR NOTES...
Page 206: ...21 6 FOR YOUR NOTES...
Page 246: ...A 20 FOR YOUR NOTES...
Page 274: ...D 8 FOR YOUR NOTES...
Page 285: ...G D Declarat tion of C Conform G 1 mity...
Page 286: ...P N 046 005297 00 15 0...