8
-9
8
.5.13
About the Defibrillator Synchronization
If a defibrillator is connected, a defibrillator synchronization pulse (100 ms, +5V) is outputted through the
multifunctional Connector every time when the patient monitor detects an R-wave.
WARNING
Improper use of a defibrillator may cause injury to the patient. The user should determine whether to
perform defibrillation or not according to the patient’s condition.
Before defibrillation, the user must ensure both defibrillator and monitor has passed the system test and
can be safely used jointly.
8
.6
About ST Monitoring
Mortara ST segment analysis is not intended for neonatal patients.
ST segment analysis calculates ST segment elevations and depressions for individual leads and then displays them
as numerics in the ST1 and ST2 areas.
A positive value indicates ST segment elevation; a negative value indicates ST segment depression.
Measurement unit of the ST segment: mV or mm. You can set the unit in the [
Unit Setup
] menu from the [
User
Maintenance
] menu.
Measurement range of the ST segment: -2.0 mV to +2.0 mV.
WARNING
The ST algorithm has been tested for accuracy of the ST segment data. The significance of the ST segment
changes need to be determined by a clinician.
8
.6.1
Switching ST On and Off
To switch ST monitoring on or off:
1.
In the [
ECG Setup
] menu, select [
ST Analysis >>
].
2. Select
[
ST Analysis
] to toggle between [
On
] and [
Off
].
Reliable ST monitoring can hardly be ensured if:
You are unable to get a lead that is not noisy.
Arrhythmias such as atrial fib/flutter cause irregular baseline.
The patient is continuously ventricularly paced.
The patient has left bundle branch block.
In these cases, you may consider switching ST monitoring off.
Summary of Contents for BeneView T1
Page 1: ...BeneView T1 Patient Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 20: ...12 FOR YOUR NOTES...
Page 38: ...2 14 FOR YOUR NOTES...
Page 56: ...4 8 FOR YOUR NOTES...
Page 64: ...5 8 FOR YOUR NOTES...
Page 116: ...8 30 FOR YOUR NOTES...
Page 130: ...11 8 FOR YOUR NOTES...
Page 138: ...12 8 FOR YOUR NOTES...
Page 140: ...13 2 FOR YOUR NOTES...
Page 174: ...16 12 FOR YOUR NOTES...
Page 190: ...18 10 FOR YOUR NOTES...
Page 206: ...21 6 FOR YOUR NOTES...
Page 246: ...A 20 FOR YOUR NOTES...
Page 274: ...D 8 FOR YOUR NOTES...
Page 285: ...G D Declarat tion of C Conform G 1 mity...
Page 286: ...P N 046 005297 00 15 0...