A-5
A.5.3
LEDs
Main unit
Alarm lamp
1 (two colors: yellow and red)
Power on LED
1 (green)
External power LED
1 (green)
Battery LED
1 (green)
T1 docking station
Connection status indicator
1 (green)
External power supply indicator
1 (green)
A.5.4
Audio Indicator
Speaker
Give alarm tones (45 to 85 dB), key tones, QRS tones; support PITCH TONE and
multi-level tone modulation; alarm tones comply with IEC60601-1-8.
A.5.5
Outputs
Analog Output
Standard
Meets the requirements of IEC60601-1 for short-circuit protection and leakage
current
ECG Analog Output
Bandwidth
(-3dB; reference frequency: 10Hz)
Diagnostic mode:
Monitor mode:
Surgical mode:
ST mode:
0.05 to 150 Hz
0.5 to 40 Hz
1 to 20 Hz
0.05 to 40 Hz
QRS delay
≤
25 ms (in diagnostic mode, and with Paced off )
Sensitivity
1V/mV ±5%
PACE rejection/enhancement
Pace enhancement
Signal amplitude: Voh≥2.5V
Pulse width: 10ms±5%
Signal rising and falling time: ≤100μs
IBP Analog Output
Bandwidth
(-3dB; reference frequency:1Hz)
DC to 40 Hz
Max transmission delay
30 ms
Sensitivity 1
V/100
mmHg
±5%
Defib Sync Pulse
Output impedance
≤
100Ω
Max time delay
35 ms (R-wave peak to leading edge of pulse)
Amplitude
High level: 3.5 to 5 V, providing a maximum of 10 mA output current;
Low level: < 0.5 V, receiving a maximum of 5 mA input current.
Pulse width
100 ms ±10%
Rising and falling time
≤1 ms
Summary of Contents for BeneView T1
Page 1: ...BeneView T1 Patient Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 20: ...12 FOR YOUR NOTES...
Page 38: ...2 14 FOR YOUR NOTES...
Page 56: ...4 8 FOR YOUR NOTES...
Page 64: ...5 8 FOR YOUR NOTES...
Page 116: ...8 30 FOR YOUR NOTES...
Page 130: ...11 8 FOR YOUR NOTES...
Page 138: ...12 8 FOR YOUR NOTES...
Page 140: ...13 2 FOR YOUR NOTES...
Page 174: ...16 12 FOR YOUR NOTES...
Page 190: ...18 10 FOR YOUR NOTES...
Page 206: ...21 6 FOR YOUR NOTES...
Page 246: ...A 20 FOR YOUR NOTES...
Page 274: ...D 8 FOR YOUR NOTES...
Page 285: ...G D Declarat tion of C Conform G 1 mity...
Page 286: ...P N 046 005297 00 15 0...