A-14
*Measurement accuracy verification: In adult and pediatric modes, the blood pressure measurements measured with this
device are in compliance with the Standard for Non-invasive sphygmomanometers (ISO 81060-2) in terms of mean error and
standard deviation by comparing with intra-arterial or auscultatory measurements (depending on the configuration) in a
typical patient population. For auscultatory reference, the 5th Korotkoff sound was used to determine the diastolic pressure.
In neonatal mode, the blood pressure measurements measured with this device are in compliance with the Standard for
Non-invasive sphygmomanometers (ISO 81060-2) in terms of mean error and standard deviation by comparing with
intra-arterial measurements (depending on the configuration) in a typical patient population.
A.8.6
Temp
Standards
Meet standard of ISO 80601-2-56
Technique Thermal
resistance
Operating mode
Direct mode
Measurement range
0 to 50
℃
(32 to 122
℉
)
Resolution 0.1
℃
Accuracy ±0.1
℃
(without probe)
Refreshing rate
1 s
Response time
<5 s
Minimum time for accurate
measurement
Body surface: <100 s
Body cavity: <80 s
Minimum time between
measurements
Body surface probe: <100 s
Body cavity probe: <80 s
Alarm limit
Range
Step
T1/T2 High
(low limit +1) to 50
℃
0.1
℃
T1/T2 Low
0.1 to (high limit -1)
℃
TD High
0.1 to 50
℃
Summary of Contents for BeneView T1
Page 1: ...BeneView T1 Patient Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 20: ...12 FOR YOUR NOTES...
Page 38: ...2 14 FOR YOUR NOTES...
Page 56: ...4 8 FOR YOUR NOTES...
Page 64: ...5 8 FOR YOUR NOTES...
Page 116: ...8 30 FOR YOUR NOTES...
Page 130: ...11 8 FOR YOUR NOTES...
Page 138: ...12 8 FOR YOUR NOTES...
Page 140: ...13 2 FOR YOUR NOTES...
Page 174: ...16 12 FOR YOUR NOTES...
Page 190: ...18 10 FOR YOUR NOTES...
Page 206: ...21 6 FOR YOUR NOTES...
Page 246: ...A 20 FOR YOUR NOTES...
Page 274: ...D 8 FOR YOUR NOTES...
Page 285: ...G D Declarat tion of C Conform G 1 mity...
Page 286: ...P N 046 005297 00 15 0...