C-4
Arrhythmia Alarm Settings (Mortara)
Item Name
Configurable
Default
In Config Mode
In Monitor Mode
Alarm Switch
Alarm Level
Asystole Alarm
Yes
Yes
On
High
VFib Alarm
Yes
Yes
On
High
Vtac Alarm
Yes
Yes
On
High
PVCs/min Alarm
Yes
Yes
General, OR, ICU, NICU: Off
CCU: On
Med
R on T Alarm
Yes
Yes
General, OR, ICU, NICU: Off
CCU: On
Med
Run PVCs Alarm
Yes
Yes
Off
Low
Couplet Alarm
Yes
Yes
Off
Prompt
Vent. Rhythm Alarm
Yes
Yes
General, OR, ICU, NICU: Off
CCU: On
Med
Bigeminy Alarm
Yes
Yes
General, OR, ICU, NICU: Off
CCU: On
Med
Trigeminy Alarm
Yes
Yes
General, OR, ICU, NICU: Off
CCU: On
Med
Tachy Alarm
Yes
Yes
Off
Med
Brady Alarm
Yes
Yes
Off
Med
PNP Alarm
Yes
Yes
Off
Prompt
PNC Alarm
Yes
Yes
Off
Prompt
Missed Beats Alarm
Yes
Yes
Off
Med
Multif. PVC
Yes
Yes
Off
Med
Irr. Rhythm Alarm
Yes
Yes
Off
Prompt
C.1.2
RESP
Item Name
Configurable
Default
In Config Mode
In Monitor Mode
Alarm Yes
Yes
On
Alm Lev
Yes
Yes
Med
RR High
Yes
Yes
Adu, Ped:
Neo:
30
100
RR Low
Yes
Yes
Adu, Ped:
Neo:
8
30
Apnea Delay
Yes
Yes
Adu, Ped:
Neo:
20
15
Lead Yes
Yes
II
Gain Yes
Yes
X2
Sweep Yes
Yes
6.25
mm/s
Detection Mode
Yes
Yes
Auto
RR Source
No
Yes
Auto
Summary of Contents for BeneView T1
Page 1: ...BeneView T1 Patient Monitor Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 20: ...12 FOR YOUR NOTES...
Page 38: ...2 14 FOR YOUR NOTES...
Page 56: ...4 8 FOR YOUR NOTES...
Page 64: ...5 8 FOR YOUR NOTES...
Page 116: ...8 30 FOR YOUR NOTES...
Page 130: ...11 8 FOR YOUR NOTES...
Page 138: ...12 8 FOR YOUR NOTES...
Page 140: ...13 2 FOR YOUR NOTES...
Page 174: ...16 12 FOR YOUR NOTES...
Page 190: ...18 10 FOR YOUR NOTES...
Page 206: ...21 6 FOR YOUR NOTES...
Page 246: ...A 20 FOR YOUR NOTES...
Page 274: ...D 8 FOR YOUR NOTES...
Page 285: ...G D Declarat tion of C Conform G 1 mity...
Page 286: ...P N 046 005297 00 15 0...