8-16
8.7.6
Changing QT Settings
8.7.6.1
Setting QT Alarm Properties
To set QT alarm properties,
1. Select
[
Alarm Setup>>
] from the
[
QT Analysis]
menu.
2.
Set QTc and ΔQTc alarm properties.
8.7.6.2
Selecting Leads for QT Calculation
You can select one lead or all leads for QT calculation. To do so, select [
Analysis Lead
] from the [
QT Analysis
] menu. [
All
]
is selected by default. This means all leads are used for QT calculation.
8.7.6.3
Changing the QTc Formula
The monitor uses as a default the Hodges correction formula to correct the QT interval for heart rate. To change the QTc
formula, select [
QTc Formula
] from the [
QT Analysis
] menu.
Hodges:
60
75
.
1
HeartRate
QT
QTc
Bazett:
2
1
60
HeartRate
QT
QTc
Fridericia:
3
1
60
HeartRate
QT
QTc
Framingham:
HeartRate
QT
QTc
60
1
154