17.1 General Specifications
295
Chapt
er
17
S
peci
fic
at
ions
17
Basic Specifications
Recording Section
Specifications
Chapter 17
17.1 General Specifications
Measurement functions
•
Memory Function (high-speed data saving)
•
Recorder Function (real time recording)
•
X-Y Recorder Function
•
FFT Function
No. of channels (max.)
Analog 16 ch Logic 16 channels (using the included logic probe input connector, ground
is common with the GND terminal) or Analog 10 ch Logic 64 channels (using supplied
equipment plus up to three Model 8973 Logic Units)
Memory capacity
16 channels
: 4 MW /channel,
8 channels
: 8 MW /channel
4 channels
: 16 MW /channel,
2 channels
: 32 MW /channel
Maximum sampling rate
20 MS / s (All channels simultaneously)
Timebase accuracy
± 0.01% (Relative grid timing error)
External control terminals
External Trigger Input, Trigger Output, External Sampling Input, GND, External Output 2 terminals
(GO, NG), External Input 3 terminals (START, STOP, PRINT)
Clock functions
Auto calendar, auto leap year judgment, 24-hour timer
Accuracy: ±100 ppm (within the operating temperature range) Reference value: ± 10 ppm (25°C)
Backup battery life
Approx. Ten years for clock and settings (@25°C, 77°F)
Operating environment
Indoors, Pollution degree 2, up to 2000 m (6562-ft.) ASL
Operating temperature and
humidity
-10 to 40°C (14 to 104°F), 20 to 80%RH (non-condensation)
•
When using the printer: 0 to 40°C (32 to 104°F), 20 to 80%RH (non-condensation)
•
When using the 9664 HD Unit: 5 to 40°C (41 to 104°F), 20 to 80%RH (non-condensation)
Temperature and humidity
range for guaranteed
accuracy
23 ± 5°C (73.4 ± 41°F), 20 to 80%RH (non-condensation)
Period of guaranteed accuracy
1 year
Storage temperature and hu-
midity
-20 to 50°C (-4 to 422°F), 90%RH or less (non-condensation)
Isolation resistance and with-
stand voltage
Between chassis and power line: 1.69 kVAC for 15 s at least 100 M
Ω
at 500 VDC
Between input module and chassis: 2.704 kVAC for 15s at least 100 M
Ω
at 500 VDC
Power source
Rated power supply voltage: 100 to 240 VAC (continuous input) (Voltage fluctuations of ±10%
from the rated supply voltage are taken into account.)
Rated power supply frequency: 50/60 Hz
DC power supply input: 10 to 28 VDC (compatible with Model 9784 DC Power Unit)
Overvoltage category II (anticipated transient overvoltage 2500 V)
Maximum rated power
130 VA max. / When using the printer: 220 VA max.
Dimensions
Approx. 351W × 261H × 140D mm (13.82"W × 10.28"H × 5.51"D) (sans protrusions),
Approx. 365W × 307H × 164D mm (14.37"W × 12.09"H × 6.46"D) (within protrusions)
Mass
Approx. 7.5 kg (Instrument)
Applicable standards
Safety
EN61010
EMC
EN61326 Class A, EN61000-3-2, EN61000-3-3
Recording system
Thermosensitive recording system using thermal line head
Recording paper
Approx. 216 mm × 30 m (8.50” x 98.43-ft)
roll-type thermosensitive paper (9231 Recording Paper)
Recording width
Overall recording width Approx. 208 mm ±1mm (8.19”±0.04”),
Waveform portion Approx. 200 mm ±1 mm (7.87”±0.04”)
Recording speed
Maximum 50 mm/s
Paper feeding accuracy
± 1% (25°C (77°F), 60%RH)
Recording paper loading
One-touch system
Summary of Contents for 8847
Page 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Page 2: ......
Page 18: ...Operating Precautions 8 ...
Page 26: ...1 4 Basic Key Operations 16 ...
Page 74: ...4 4 Waveform Observation 64 ...
Page 98: ...5 5 Managing Files 88 ...
Page 110: ...6 6 Miscellaneous Printing Functions 100 ...
Page 126: ...7 8 Seeing Block Waveforms 116 ...
Page 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Page 198: ...10 6 Numerical Calculation Type and Description 188 ...
Page 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Page 304: ...16 2 External I O 294 ...
Page 362: ...Appendix 5 FFT Definitions A28 ...
Page 368: ...Index Index 6 ...
Page 369: ......
Page 370: ......
Page 371: ......