9.3 Triggering by Analog Signals
154
The steps for making settings and selecting the type of analog trigger are described below.
The Trigger settings window (
[Analog Trg.]
sheet) is used.
9.3
Triggering by Analog Signals
9.3.1 Analog Trigger Settings and Types
Procedure
To open the screen: Press the
DISP
key
→
Waveform screen
→
Press the
TRIG.SET
key
→
Trigger settings
window (
[Analog Trg.]
sheet)
1
Move the cursor to the
[Type]
item of the channel for which to make the setting.
2
Select the trigger type with the
F
keys.
3
Move the cursor to the parameter item with the
CURSOR
keys.
4
Select the parameter value with the
F
keys.
1
3
1. Level Trigger (
2. In-Window Trigger
Out-of-Window Trigger (
3. Voltage Sag Trigger (
4. In-Period Trigger
Out-of-Period Trigger (
5. Glitch Trigger (
To copy the setting to another channel
The Trigger settings window (
[Analog Trig]
sheet) can be used to copy a setting.
See:
"8.9 Copying settings to other channels (calculation No.) (Copy function)" (
When the FFT function is used and
[Reference]
item on the
[Status]
sheet is
set to
[From Memory]
, an analog trigger cannot be set.
Summary of Contents for 8847
Page 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Page 2: ......
Page 18: ...Operating Precautions 8 ...
Page 26: ...1 4 Basic Key Operations 16 ...
Page 74: ...4 4 Waveform Observation 64 ...
Page 98: ...5 5 Managing Files 88 ...
Page 110: ...6 6 Miscellaneous Printing Functions 100 ...
Page 126: ...7 8 Seeing Block Waveforms 116 ...
Page 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Page 198: ...10 6 Numerical Calculation Type and Description 188 ...
Page 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Page 304: ...16 2 External I O 294 ...
Page 362: ...Appendix 5 FFT Definitions A28 ...
Page 368: ...Index Index 6 ...
Page 369: ......
Page 370: ......
Page 371: ......