13.3 Setting FFT Analysis Conditions
221
13
Chap
ter
13 FFT
Fun
cti
on
3
When [Parameter] setting contents are displayed
Set the parameter.
Move the cursor to the
[Parameter]
column of the Analysis No.
to set.
Select
4
Select the channel for analysis.
Move the cursor to the
[Ch1]
item.
Select which channel number to use.
5
Set the horizontal and vertical axes displaying the calcu-
lation results.
Move the cursor to the
[X Axis]
or
[Y Axis]
.
Set the contents of the calculation results to be displayed on
the horizontal and vertical axes.
(Selectable display contents vary by analysis mode.)
See:
"Analysis Modes and X/Y Axis Display" (
Y-axis display
X-axis display
Analyze
Parameter
Setting Contents
1/1 Octave,
1/3 Octave
Filter: Normal
nables the octave filter.
See:
Filter: Sharp
Phase Spectrum
1ch FFT
Calculates the phase of
[Ch1]
.
2ch FFT
Calculates the phase difference between
[Ch1]
and
[Ch2]
.
LPC (dencity)
Order :2 to 64
Larger numerical values make finer spectrum compo-
nents visible.
Lin-Mag
Analysis results are displayed as amplitude values.
Log-Mag
Analysis results are displayed as dB values.
The dB reference is 1 eu. (As a voltage example, 1 V is
0 dB.)
Lin-Real
The real-number component of analysis results are dis-
played.
Lin-Imag
The imaginary component of analysis results are dis-
played.
Linear
Frequency-axis is displayed linearly.
Log
Frequency-axis is displayed logarithmically.
This is convenient when the data of interest is at the lower end
of the frequency range, such as for sound and vibration.
Analysis channel setting
For any of the following analysis modes,
set both channels 1 and 2.
Transfer Function, Impulse Response,
Cross-correlation Function, Cross Power
Spectrum, Coherence Function, Phase
Spectrum (2ch FFT)
To analyze using external sampling
The horizontal axis ( x-axis) displays the
number of data points.
How do I copy settings to other calculation No.?
See:
"8.9 Copying settings to other channels (calculation No.) (Copy function)" (
Summary of Contents for 8847
Page 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Page 2: ......
Page 18: ...Operating Precautions 8 ...
Page 26: ...1 4 Basic Key Operations 16 ...
Page 74: ...4 4 Waveform Observation 64 ...
Page 98: ...5 5 Managing Files 88 ...
Page 110: ...6 6 Miscellaneous Printing Functions 100 ...
Page 126: ...7 8 Seeing Block Waveforms 116 ...
Page 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Page 198: ...10 6 Numerical Calculation Type and Description 188 ...
Page 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Page 304: ...16 2 External I O 294 ...
Page 362: ...Appendix 5 FFT Definitions A28 ...
Page 368: ...Index Index 6 ...
Page 369: ......
Page 370: ......