16.2 External I/O
289
13
Chapt
er 16
Ext
e
rn
al Cont
rol
16
External control signals can be applied to start and stop measurement, and to print and save data.
The factory-default settings are to
[Start]
,
[Stop]
, and
[Print]
.
1.
Connect the cables for the corresponding external input signals to START/
EXT.IN1, STOP/EXT.IN2, PRINT/EXT.IN3, and GND terminals.
See:
"16.1 Connecting External Control Terminals" (
2.
Move the cursor to the Press the
SYSTEM
key to open the Environment sheet,
and move the cursor to
[START/EXT IN1]
,
[STOP/EXT IN2]
, or
[PRINT/EXT
IN3]
.
3.
Select the operation performed by the instrument in response to external signal input.
Select
4.
Short circuit the terminal and GND, or input a HIGH level (3.0 to 5.0 V) or LOW
level (0 to 0.8 V) pulse wave or rectangular wave to the terminal.
Control with the LOW level of the input waveform.
16.2 External I/O
16.2.1 External Input (START/EXT.IN1) (STOP/EXT.IN2)
(PRINT/EXT.IN3)
Signal Input Procedure
Start
Start measurement. (Is not affected by
[START Action]
Stop
Stop measurement. (Operations after measurement, such as numerical calcula-
tions and automatic saving will be carried out)
Start/Stop
Start measurement on LOW level, and stop measurement on HIGH level.
Abort
Stop measurement immediately. (Operations after measurement, such as nu-
merical calculations and automatic saving will not be carried out)
Same operation as
key (Not valid during Selective Printing (
Save
Save to the media specified for the
SAVE
key, according to the specified condi-
tions. (Not valid during Selective Saving (
Pen Up/Down
A LOW level signals 'pen-down' operation and a HIGH level signals 'pen-up' op-
eration (for X-Y recording).
• For STOP operation, follow
[STOP Action]
• External input is not available when the HELP screen or a dialog window is
being displayed.
Voltage range
HIGH level: 3.0 to 5.0 V , LOW level: 0 to 0.8 V
Pulse width
HIGH level: 20 ms or greater, LOW level: 30 ms or greater
Maximum input
voltage
-0.5 to 7 V
START/EXT.IN1
STOP/EXT.IN2
PRINT/EXT.IN3
470
Ω
GND
2200 pF
5 V
3.3 k
Ω
HIGH
3.0 to 5.0 V
LOW
0 to 0.8 V
30 ms or greater
20 ms or
greater
5 V
Summary of Contents for 8847
Page 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Page 2: ......
Page 18: ...Operating Precautions 8 ...
Page 26: ...1 4 Basic Key Operations 16 ...
Page 74: ...4 4 Waveform Observation 64 ...
Page 98: ...5 5 Managing Files 88 ...
Page 110: ...6 6 Miscellaneous Printing Functions 100 ...
Page 126: ...7 8 Seeing Block Waveforms 116 ...
Page 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Page 198: ...10 6 Numerical Calculation Type and Description 188 ...
Page 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Page 304: ...16 2 External I O 294 ...
Page 362: ...Appendix 5 FFT Definitions A28 ...
Page 368: ...Index Index 6 ...
Page 369: ......
Page 370: ......
Page 371: ......