18.3 Error Messages
315
Chapte
r 18
Main
tenanc
e and
Ser
vi
ce
14
18
A screen message appears whenever an error occurs. In either case, take the remedial action indicated.
A beep may sound if the beeper setting on the
[Environment]
sheet is
[Warn]
or
[Warn+Action]
.
See:
"Chapter 14 System Environment Settings" (
18.3 Error Messages
Message
Displayed Message
Number (Msg No.)
Warning Display
Appears just once when an error
occurs. Disappears within a few
seconds.
Also disappears when any key is
pressed.
Displayed Warnings
Msg
No.
Message
Remedial Action
Reference
3
Out of paper.
Load more paper.
"2.4 Loading Recording Pa-
per" (
4
Printer lever is opened.
Close the printer lever.
5
Recording length is set to Continuous.
When the Recording Length (Shot) is set to
Continuous (Cont.), real-time printing is not
available with fast timebase settings.
"3.4.3 Recording Length
(number of divisions)" (
6
Cannot set.(Time/Div 10ms-200ms).
When the Recording Length (Shot) is set to
Continuous (Cont.), the printer cannot be
used.
10
Mount media.
Set the recording media.
"2.3 Recording Media Prep-
aration" (
11
Illegal format.
Wrong recording format. Reformat the media.
"2.3.2 Formatting Storage
Media" (
12
Can not write.
Write-protection is enabled on the storage
media. Disable it.
-
13
Disk full while accessing the file.
Saving is not possible because of insufficient
space on the storage media. Delete files or re-
place the storage media. If measuring,
Stop measurement, then replace the storage
media.
"5.5.4 Deleting Files & Fold-
ers" (
14
File is read only.
File is read only. Deletion is not permitted.
-
15
Access to the file was denied.
An internal fault may have occurred in the in-
strument. Turn the instrument off and back on
-
16
The file name already exists: cannot
save.
Change the file name.
"5.5.6 Renaming Files &
Folders" (
17
The folder name already exists.
Change the folder name.
18
Not enough free space in the folder to
create file.
Either delete files in the saving destination
folder, or change to another saving destina-
tion.
-
19
Folder not empty.
The folder is not empty, be sure you want to
delete.
-
20
The maximum length of a file name, in-
cluding its path, is 255 characters.
File names including the path must not exceed
255 characters.
-
21
Internal error.
An internal error occurred. Check the storage
media.
-
22
No waveform data to save.
Acquire waveform data.
-
24
There is no calculation result.
There is no calculation result. Print results af-
ter performing calculation.
"Chapter 10 Numerical Cal-
culation Functions" (
25
Can not select this media.
Automatic saving can only be specified in CF/
HDD.
-
26
Invalid folder path.
Route cannot be specified.
-
Summary of Contents for 8847
Page 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Page 2: ......
Page 18: ...Operating Precautions 8 ...
Page 26: ...1 4 Basic Key Operations 16 ...
Page 74: ...4 4 Waveform Observation 64 ...
Page 98: ...5 5 Managing Files 88 ...
Page 110: ...6 6 Miscellaneous Printing Functions 100 ...
Page 126: ...7 8 Seeing Block Waveforms 116 ...
Page 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Page 198: ...10 6 Numerical Calculation Type and Description 188 ...
Page 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Page 304: ...16 2 External I O 294 ...
Page 362: ...Appendix 5 FFT Definitions A28 ...
Page 368: ...Index Index 6 ...
Page 369: ......
Page 370: ......
Page 371: ......