18.1 Trouble Shooting
312
18.1 Trouble Shooting
If damage is suspected, check the "Troubleshooting" section before
contacting your dealer or Hioki representative.
If Power and Operating Keys Malfunction
Symptom
Check Item, or Cause
Remedy and Reference
The display does not
appear when you turn the
power on.
Is the power cord disconnected?
Are connections made correctly?
Verify that the power cord is connected
properly.
Keys do not work.
• Is any key being held down?
• Verify key operation.
• Is the key-lock state active (Key-Lock
message displayed)?
• Deactivate the key-lock state. (Hold the
CURSOR
keys for three seconds)
If the Display or Operations Malfunction
Symptom
Check Item, or Cause
Remedy and Reference
Blank screen
Is the backlight saver active? Press any
key to check.
Turn the backlight saver off.
A waveform does not
appear when you press
the
START
key.
• Is the “Pre-Trigger wait” message
displayed?
• Is the "Trigger wait" message dis-
played?
When pre-triggering is enabled, triggering is
ignored until the pre-trigger portion of the
waveform has been acquired. Recording
starts when a trigger occurs.
No changes occur in the
displayed waveform.
• Is the clamp sensor or connection
cable connected correctly?
• Is the vertical axis (voltage axis)
range set properly?
• Is the low-pass filter enabled?
Verify that the clamp sensor or connection
cable is connected correctly.
Verify the input channel settings.
While measuring with the
memory function, the dis-
played frequency is much
lower than the actual fre-
quency.
Aliasing may be occurring.
Change the timebase to use a faster sam-
pling rate.
"3.4.2 Time Axis Range and Sampling
Rate" (
Some channels are unus-
able.
Is the channel enabled for use?
"8.4 Setting Channels to Use (Extending
the Recording Length)" (
The displayed waveform
size does not change
when the input range is
changed.
Is the Variable function enabled?
Turn the Variable function off.
"8.6 Variable Function (Setting the
Waveform Display Freely)" (
Summary of Contents for 8847
Page 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Page 2: ......
Page 18: ...Operating Precautions 8 ...
Page 26: ...1 4 Basic Key Operations 16 ...
Page 74: ...4 4 Waveform Observation 64 ...
Page 98: ...5 5 Managing Files 88 ...
Page 110: ...6 6 Miscellaneous Printing Functions 100 ...
Page 126: ...7 8 Seeing Block Waveforms 116 ...
Page 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Page 198: ...10 6 Numerical Calculation Type and Description 188 ...
Page 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Page 304: ...16 2 External I O 294 ...
Page 362: ...Appendix 5 FFT Definitions A28 ...
Page 368: ...Index Index 6 ...
Page 369: ......
Page 370: ......
Page 371: ......