13.3 Setting FFT Analysis Conditions
210
F
Select the data to be used for FFT analysis.
There are two analysis methods: analysis using new measurements and analysis of data measured
using the memory function.
13.3.2 Selecting the Data Source for Analysis
1
Select the input data source.
Move the cursor to the
[Reference]
item.
Select
2
When finished making settings, press the
START
key
For the [New Data] case
Measurement starts to acquire data for the number of analysis points
specified as the
[Sampling Point]
, and FFT analysis is performed.
For the [From Memory] case
Analysis is performed on the number of specified points from data
previously recorded in memory (Memory function data) .
The analysis starting point can also be specified.
See:
"13.8.1 Analyzing after Specifying an Analysis Starting Point" (
The frequency range is selected automatically.
See:
"Relationship Between Frequency Range, Resolution and Number of
Analysis Points" (
New Data
Acquire a new waveform for analysis.
From Memory
Calculates data measured using the memory function.
Procedure
To open the screen: Press the
STATUS
key
→
[Status]
sheet
See:
To set from the Waveform screen (
When the input data [Reference]
is
[From Memory]
Analysis is performed until the specified
number of FFT analysis points have been
processed, then the data is shifted by that
amount and analysis repeats until all of the
previously acquired data has been pro-
cessed. (If the amount of data is less than
the specified number of FFT analysis
points, no analysis occurs.)
When no trace is displayed after press-
ing the START key
Analysis is impossible if
[From Memory]
is selected as the input data source and
no recorded data exists in the instrument's
memory.
Either select
[New Data]
as the input data
source, or load the data to be analyzed
before pressing the
START
key again.
1
When the
[Reference]
is set to
[From Memory]
, the frequency is automatically
set.
The setting cannot be changed.
Summary of Contents for 8847
Page 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Page 2: ......
Page 18: ...Operating Precautions 8 ...
Page 26: ...1 4 Basic Key Operations 16 ...
Page 74: ...4 4 Waveform Observation 64 ...
Page 98: ...5 5 Managing Files 88 ...
Page 110: ...6 6 Miscellaneous Printing Functions 100 ...
Page 126: ...7 8 Seeing Block Waveforms 116 ...
Page 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Page 198: ...10 6 Numerical Calculation Type and Description 188 ...
Page 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Page 304: ...16 2 External I O 294 ...
Page 362: ...Appendix 5 FFT Definitions A28 ...
Page 368: ...Index Index 6 ...
Page 369: ......
Page 370: ......
Page 371: ......