17.6 Input Modules Specifications
305
Chapt
er
17
S
peci
fic
at
ions
17
17.6.2 8967 Temp Unit
Temperature and humidity range
for guaranteed accuracy
23 ± 5°C (73 ± 9F°), 20 to 80% RH (when zero adjustment is executed 30 minutes after power on)
Period of guaranteed accuracy
1 year
No. of input channels
2 channels
Input terminals
Pushbutton-type terminal block (2 terminals/ch)
Sensor
Thermocouple (K, E, J, T, N, W, R, S, B)
Measurement ranges
Measurable range
Resolution
Measurement accuracy (f.s.=20
div)
Reference junction compensa-
tion accuracy
± 1.5°C (Reference junction compensation: When internal, add to thermocouple measurement ac-
curacy)
Reference junction compensa-
tion
Selectable internal or external (during thermocouple measurement)
Temperature characteristic
Add (meas. accuracy × 0.1)°C to meas. accuracy
Data refresh
Selectable from normal, fast or slow
Open-circuit detection
Selectable On or Off
Input resistance
100M
Ω
(with open-circuit detection Off) or 5.1 M
Ω
± 5% with open-circuit detection On)
Common mode rejection ratio
80 dB minimum (at 50/60 Hz, with signal source
100
Ω
, and Fast data refresh setting)
100 dB minimum (at 50/60Hz, with signal source
100
Ω
, and Normal data refresh setting)
Input type
Unbalanced (floating)
Maximum rated voltage to earth
300 VAC/DC (between input channels and chassis, and between input channels)
Measurement category II, (anticipated transient overvoltage 2500 V)
Dielectic strength
2.704 kVAC for 15 seconds (between input channels and chassis, and between input channels)
Operating temperature and hu-
midity
Same as the host Memory HiCorder
Storage temperature and humidity
Temperature-20 to 50°C, Humidity90% RH or less (non-condensating)
Operating environment
Same as the host Memory HiCorder
Dimensions
Approx. 106 mmW × 19.8 mmH × 204.5mmD (4.17"W × 0.78"H × 8.05"D)
Mass
Approx. 240 g (8.5 oz)
Option
Ferrite clamp-on choke........................................................................................ 2
Effect of radiated ratio-frequency
electromagnetic field
± 2% f.s. (max) at 3 V/m
Effect of conducted radio- fre-
quency electromagnetic field
± 2% f.s. (max) at 3 V
Applicable standards
Safety
EN61010
EMC
EN61326 Class A
Sensor
Range
Measurable range
Resolu-
tion
Measurement accuracy
Thermocouple
(not including
standard junction
compensation
accuracy)
K
*1
10°C/div
-100°C to 200°C
0.01°C
±0.1% f.s. ± 1°C
± 0.1% f.s. ± 2°C
(-200°C to 0°C)
50°C/div
-200°C to 1000°C
0.05°C
100°C/div
-200°C to 1350°C
0.1°C
J
*1
10°C/div
-100°C to 200°C
0.01°C
50°C/div
-200°C to 1000°C
0.05°C
100°C/div
-200°C to 1100°C
0.1°C
E
*1
10°C/div
-100°C to 200°C
0.01°C
50°C/div
-200°C to 800°C
0.05°C
100°C/div
-200°C to 800°C
0.1°C
T
*1
10°C/div
-100°C to 200°C
0.01°C
50°C/div
-200°C to 400°C
0.05°C
100°C/div
-200°C to 400°C
0.1°C
N
*1
10°C/div
-100°C to 200°C
0.01°C
50°C/div
-200°C to 1000°C
0.05°C
100°C/div
-200°C to 1300°C
0.1°C
R
*1
10°C/div
0°C to 200°C
0.01°C
± 0.1% f.s. ± 3.5°C (0°C to
400°C)
(except B, for which accu-
racy is not guaranteed be-
low 400°C)
± 0.1% f.s. ± 3°C (400°C
and higher)
50°C/div
0°C to 1000°C
0.05°C
100°C/div
0°C to 1700°C
0.1°C
S
*1
10°C/div
0°C to 200°C
0.01°C
50°C/div
0°C to 1000°C
0.05°C
100°C/div
0°C to 1700°C
0.1°C
B
*1
50°C/div
400°C to 1000°C
0.05°C
100°C/div
400°C to 1800°C
0.1°C
W
*2
(WRe5-26)
10°C/div
0°C to 200°C
0.01°C
50°C/div
0°C to 1000°C
0.05°C
100°C/div
0°C to 2000°C
0.1°C
*1: JIS C 1602-1995, *2: ASTM E-988-96
Selection
Fast
Normal
Slow
Data Refresh Rate
1.2 ms
100 ms
500 ms
Summary of Contents for 8847
Page 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Page 2: ......
Page 18: ...Operating Precautions 8 ...
Page 26: ...1 4 Basic Key Operations 16 ...
Page 74: ...4 4 Waveform Observation 64 ...
Page 98: ...5 5 Managing Files 88 ...
Page 110: ...6 6 Miscellaneous Printing Functions 100 ...
Page 126: ...7 8 Seeing Block Waveforms 116 ...
Page 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Page 198: ...10 6 Numerical Calculation Type and Description 188 ...
Page 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Page 304: ...16 2 External I O 294 ...
Page 362: ...Appendix 5 FFT Definitions A28 ...
Page 368: ...Index Index 6 ...
Page 369: ......
Page 370: ......
Page 371: ......