9.6 Applying an External Trigger (External Trigger)
165
9
Chapt
er
9
T
rigger
Set
tings
An external signal applied to the External Control terminal can serve as a trigger source. It can also
be used to synchronously drive parallel triggering of multiple instruments.
9.6
Applying an External Trigger (External Trig-
ger)
Procedure
To open the screen: Press the
DISP
key
→
Waveform screen
→
Press the
TRIG.SET
key
→
Trigger settings window
1
Enable external triggering.
Move the cursor to the
[Ext Trig]
item.
Select
2
Apply the input signal to the external trigger (EXT.TRIG)
terminal.
See:
"16.2.5 External Trigger terminal (EXT.TRIG)" (
Off
Disables external triggering.(default setting)
On
Enables external triggering.
9.7
Triggering Manually (Manual Trigger)
Triggers can be applied manually by pressing Manual trigger key. Manual trig-
gering takes priority over all other trigger sources, regardless of settings.
To stop recording:
Press the
STOP
key.
One
keypress
Recording is carried out for one recording length and then
stops.
Two
keypresses
Recording stops when key is pressed.
As for other trigger types, triggering does not occur during pre-trigger standby.
Set the trigger priority function to
[On]
.
See:
Summary of Contents for 8847
Page 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Page 2: ......
Page 18: ...Operating Precautions 8 ...
Page 26: ...1 4 Basic Key Operations 16 ...
Page 74: ...4 4 Waveform Observation 64 ...
Page 98: ...5 5 Managing Files 88 ...
Page 110: ...6 6 Miscellaneous Printing Functions 100 ...
Page 126: ...7 8 Seeing Block Waveforms 116 ...
Page 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Page 198: ...10 6 Numerical Calculation Type and Description 188 ...
Page 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Page 304: ...16 2 External I O 294 ...
Page 362: ...Appendix 5 FFT Definitions A28 ...
Page 368: ...Index Index 6 ...
Page 369: ......
Page 370: ......
Page 371: ......