13.3 Setting FFT Analysis Conditions
212
Relationship Between Frequency Range, Resolution and Number of
Analysis Points
Range
[Hz]
Sampling
frequency
[Hz]
Timebase
[/div]
(MEM)
Sampling
period
Number of FFT Analysis Points
1,000
2,000
5,000
10,000
Resolu-
tion [Hz]
Acquisi-
tion
interval
Resolu-
tion [Hz]
Acquisi-
tion
interval
Resolu-
tion [Hz]
Acquisi-
tion
interval
Resolu-
tion [Hz]
Acquisi-
tion
interval
8 M *
1
20 M
5
μ
s
50 ns
20 k
50
μ
s
10 k
100
μ
s
4 k
250
μ
s
2 k
500
μ
s
4 M *
1
10 M
10
μ
s
100 ns
10 k
100
μ
s
5 k
200
μ
s
2 k
500
μ
s
1 k
1 ms
2 M *
1
5 M
20
μ
s
200 ns
5 k
200
μ
s
2.5 k
400
μ
s
1 k
1 ms
500
2 ms
800 k *
1
2 M
50
μ
s
500 ns
2 k
500
μ
s
1 k
1 ms
400
2.5 ms
200
5 ms
400 k *
1
1 M
100
μ
s
1
μ
s
1 k
1 ms
500
2 ms
200
5 ms
100
10 ms
200 k *
1
500 k
200
μ
s
2
μ
s
500
2 ms
250
4 ms
100
10 ms
50
20 ms
80 k *
1
200 k
500
μ
s
5
μ
s
200
5 ms
100
10 ms
40
25 ms
20
50 ms
40 k
100 k
1 ms
10
μ
s
100
10 ms
50
20 ms
20
50 ms
10
100 ms
20 k
50 k
2 ms
20
μ
s
50
20 ms
25
50 ms
10
100 ms
5
200 ms
8 k
20 k
5 ms
50
μ
s
20
50 ms
10
100 ms
4
250 ms
2
500 ms
4 k
10 k
10 ms
100
μ
s
10
100 ms
5
200 ms
2
500 ms
1
1 s
2 k
5 k
20 ms
200
μ
s
5
200 ms
2.5
400 ms
1
250 ms
500 m
2 s
800
2 k
50 ms
500
μ
s
2
500 ms
1
1 s
400 m
2.5 s
200 m
5 s
400
1 k
100 ms
1 ms
1
1 s
500 m
2 s
200 m
5 s
100 m
10 s
200
500
200 ms
2 ms
500 m
2 s
250 m
4 s
100 m
10 s
50 m
20 s
80
200
500 ms
5 ms
200 m
5 s
100 m
10 s
40 m
25 s
20 m
50 s
40
100
1 s
10 ms
100 m
10 s
50 m
20 s
20 m
50 s
10 m
100 s
20
50
2 s
20 ms
50 m
20 s
25 m
40 s
10 m
100 s
5 m
200 s
8 *
2
20
5 s
50 ms
20 m
50 s
10 m
100 s
4 m
250 s
2 m
500s
4 *
2
10
10 s
100 ms
10 m
100 s
5 m
200s
2 m
500 s
1 m
1 ks
1.33 *
2
3.33
30 s
300 ms
3.33 m
300 s
1.66 m
600s
666
μ
1.5 ks
333
μ
3 ks
800 m *
2
2
50 s
500 ms
2 m
500 s
1 m
1 ks
400
μ
2.5 ks
200
μ
5 ks
667 m *
2
1.67
60 s
600 ms
1.66 m
600 s
833
μ
1.2 ks
333
μ
3 ks
166
μ
6 ks
400 m *
2
1
100 s
1 s
1 m
1 ks
500
μ
2 ks
200
μ
5 ks
100
μ
10 ks
333 m *
2
833 m
120 s
1.2 s
833
μ
1.2 ks
416
μ
2.4 ks
166
μ
6 ks
83.3
μ
12 ks
133 m *
2
333 m
300 s
3 s
333
μ
3 ks
166
μ
6 ks
66.6
μ
15 ks
33.3
μ
30 ks
The cut-off frequency of the anti-aliasing filter is the same as the frequency range.
*1. The anti-aliasing filter is turned off.
*2. Cut-off frequency is 20 Hz.
Summary of Contents for 8847
Page 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Page 2: ......
Page 18: ...Operating Precautions 8 ...
Page 26: ...1 4 Basic Key Operations 16 ...
Page 74: ...4 4 Waveform Observation 64 ...
Page 98: ...5 5 Managing Files 88 ...
Page 110: ...6 6 Miscellaneous Printing Functions 100 ...
Page 126: ...7 8 Seeing Block Waveforms 116 ...
Page 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Page 198: ...10 6 Numerical Calculation Type and Description 188 ...
Page 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Page 304: ...16 2 External I O 294 ...
Page 362: ...Appendix 5 FFT Definitions A28 ...
Page 368: ...Index Index 6 ...
Page 369: ......
Page 370: ......
Page 371: ......