8.10 Making Detailed Settings for Input Modules
147
8
Chapt
er 8
Ut
ilit
y Funct
ions
Level
This is enabled only when
[Mode]
is
[Pulse Width]
or
[Duty]
.
In pulse width/ duty rate measurements, set which level is detected when the
threshold is exceeded.
Select
See:
Opening the
[Each Ch]
sheet, Making a Channel Selection (
Smoothing
This is enabled only when
[Mode]
is
[Frequency]
or
[RPM]
.
Set smoothing.
Select
Hold
This is enabled only when
[Mode]
is
[Frequency]
or
[RPM]
.
Set hold for the frequency and summing.
Select
Selections
Description
High
Measures above the threshold value. (default setting)
Low
Measures below the threshold value.
Selections
Description
Off
The measured data is recorded as is. (A step waveform.)
(default setting)
On
The measured data is interpolated so that the waveform is smooth
and then it is output. (Further delayed from upper limit 10 kHz and
Off.)
Selections
Description
Off
(1Hz/0.5Hz/
0.2Hz/0.1Hz)
When the frequency in the brackets is reached but not determined,
it is defined as a stopped item and the measurement value be-
comes 0 Hz (0 rpm). (default setting)
On
The value defined last time is retained.
Summary of Contents for 8847
Page 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Page 2: ......
Page 18: ...Operating Precautions 8 ...
Page 26: ...1 4 Basic Key Operations 16 ...
Page 74: ...4 4 Waveform Observation 64 ...
Page 98: ...5 5 Managing Files 88 ...
Page 110: ...6 6 Miscellaneous Printing Functions 100 ...
Page 126: ...7 8 Seeing Block Waveforms 116 ...
Page 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Page 198: ...10 6 Numerical Calculation Type and Description 188 ...
Page 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Page 304: ...16 2 External I O 294 ...
Page 362: ...Appendix 5 FFT Definitions A28 ...
Page 368: ...Index Index 6 ...
Page 369: ......
Page 370: ......
Page 371: ......