8.10 Making Detailed Settings for Input Modules
149
8
Chapt
er 8
Ut
ilit
y Funct
ions
See:
Opening the
[Each Ch]
sheet, Making a Channel Selection (
Mode
Switches between voltage measurement and RMS measurement.
Select
Response
Response can be set to three speeds: Fast, Normal and Slow.
Normally set to
[Fast]
, this can be changed to
[Normal]
or
[Slow]
to stabilize the
display when measuring low frequencies, or when severe fluctuations are
present.
Select
8.10.7 Settings for the 8972 DC/RMS Unit
Selections
Description
DC
Voltage measurement (default setting)
RMS
RMS measurement
Selections
Description
Fast
Sets the response time to about 100 ms. (default setting)
Normal
Sets the response time to about 800 ms.
Slow
Sets the response time to about 5 s.
Summary of Contents for 8847
Page 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Page 2: ......
Page 18: ...Operating Precautions 8 ...
Page 26: ...1 4 Basic Key Operations 16 ...
Page 74: ...4 4 Waveform Observation 64 ...
Page 98: ...5 5 Managing Files 88 ...
Page 110: ...6 6 Miscellaneous Printing Functions 100 ...
Page 126: ...7 8 Seeing Block Waveforms 116 ...
Page 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Page 198: ...10 6 Numerical Calculation Type and Description 188 ...
Page 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Page 304: ...16 2 External I O 294 ...
Page 362: ...Appendix 5 FFT Definitions A28 ...
Page 368: ...Index Index 6 ...
Page 369: ......
Page 370: ......
Page 371: ......