9.5 Trigger by Timer or Time Intervals (Timer Trigger)
164
When a trigger is applied from a trigger source other than a timer trigger
Trigger sources set to On are all enabled.
However, trigger timing depends on the trigger source settings.
• When trigger criteria are ORed together (Trigger Source: OR)
Depending on the other trigger sources, triggering can occur before the speci-
fied trigger Start time, after the specified Stop time, or outside of the specified
Interval.
• When trigger criteria are ANDed together (Trigger Source: AND)
Triggering occurs between the specified Start and Stop times when criteria for
all trigger sources set within the specified interval are satisfied.
If the interval is set to zero, triggering occurs when criteria for all trigger
sources set between specified Start and Stop times are satisfied.
Example: measuring when both timer trigger and level triggers (Slope:
↑
) are
enabled.
When trigger criteria are ORed together (Trigger Source: OR)
Start Measurement
Start Time
Stop Time
Level Trigger
Timer Trigger
Recording
T
T
T
T
T
T
T
T
T
T
Interval
Ignored because they occur while recording
When trigger criteria are ANDed together (Trigger Source: AND)
Start Measurement
Start Time
Stop Time
Recording
T
Interval
T
T
Outside of Timer Time
Timer Trigger is not
applied, so ignored
Trigger was applied only once
within the interval, so ignored
Outside of
Timer Time
Measurement ends
at Stop time
Level Trigger
Timer Trigger
Ignored because they occur while recording
Summary of Contents for 8847
Page 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Page 2: ......
Page 18: ...Operating Precautions 8 ...
Page 26: ...1 4 Basic Key Operations 16 ...
Page 74: ...4 4 Waveform Observation 64 ...
Page 98: ...5 5 Managing Files 88 ...
Page 110: ...6 6 Miscellaneous Printing Functions 100 ...
Page 126: ...7 8 Seeing Block Waveforms 116 ...
Page 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Page 198: ...10 6 Numerical Calculation Type and Description 188 ...
Page 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Page 304: ...16 2 External I O 294 ...
Page 362: ...Appendix 5 FFT Definitions A28 ...
Page 368: ...Index Index 6 ...
Page 369: ......
Page 370: ......
Page 371: ......