8.5 Converting Input Values (Scaling Function)
130
The 9018-50 Clamp-On Probe provides 0.2 V output when measuring 10 A. So
Scaling should be set to display 10 A with 0.2 V input (and 0 A with 0 V input).
Selecting a Clamp Type ___________________________________________
1. Move the cursor to the
[Clamp]
item, and select
[Select]
.
The cursor moves to the
[Model]
item.
2. Select
[9000
]
.
The cursor moves to the
[Clamp]
.
3. Select
[9018-50]
from the clamp list with F key, and select
[Confirm]
.
Units, scaling method, and ratio are set automatically.
4. Select the same range of the clamp when using the range selection type.
Select
[10A]
here.
To reset Scaling settings:
Move the cursor to the
[Method]
, and select
[Reset]
.
To copy the scaling setting to another channel
The Channel screen -
[Scaling]
sheet can be used to copy a setting.
See:
"8.9 Copying settings to other channels (calculation No.) (Copy function)" (
Using the Scaling and Variable functions (
p.134) in combination:
The full span of output from a sensor can be displayed. (
At factory shipping, automatic correction of the variable function (
p.257) is set
to
[On]
.
At this time, the Variable setting is altered so that it is linked to (dependent upon)
the vertical axis (voltage axis) range and scaling settings. If you want the Vari-
able function setting to take priority, use either of the following procedures:
• Set Scaling first, and then set the Variable function
• Set a Variable value before Scaling, and then set Scaling.
When automatic correction of the Variable function (Variable Auto Adjustment) is
disabled (Off), the Scaling and Variable settings are unlinked (independent of
one another).
8.5.1 Scaling Setting Examples
Using a Clamp-On Probe
Example 1
Measure with the 10 A range of the Model 9018-10 Clamp-On Probe and
display the measured data in units of [A] (Amperes)
Setting Items
Setting Choice
Disp
NUM or SCI
Clamp
9018-50
Unit
*
A
Method
*
Ratio
Ratio
*
50.000
*:
Set automatically when clamp is selected.
Summary of Contents for 8847
Page 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Page 2: ......
Page 18: ...Operating Precautions 8 ...
Page 26: ...1 4 Basic Key Operations 16 ...
Page 74: ...4 4 Waveform Observation 64 ...
Page 98: ...5 5 Managing Files 88 ...
Page 110: ...6 6 Miscellaneous Printing Functions 100 ...
Page 126: ...7 8 Seeing Block Waveforms 116 ...
Page 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Page 198: ...10 6 Numerical Calculation Type and Description 188 ...
Page 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Page 304: ...16 2 External I O 294 ...
Page 362: ...Appendix 5 FFT Definitions A28 ...
Page 368: ...Index Index 6 ...
Page 369: ......
Page 370: ......
Page 371: ......