5.2 Saving Data
76
6
Making advanced settings
Available settings will differ, depending on the save type. Refer
to the table below.
7
(For memory function)
When using the memory division function
Select the blocks to be saved
Move the cursor to the
[Save Block]
.
Select
This completes the data save settings.
When the
SAVE
key is pressed, the saving operation will use
these settings.
Save Type
Settings
Description
Setting
-
-
Wave
Binary
Division
(Off, 16 M, 32 M)
Select this to divide a large file into several files for
saving.
A folder with the specified name is created and data
are saved as multiple files in that folder.
Wave Text
Save Thin
(Off, 1/2 to 1/1000)
Select this to use thinning (sub-sampling) when sav-
ing data.
Only a specified subset (one out of several data) is re-
tained.
Screen
Copy
(Hard
copy of
the
display)
Image Color
(Color, Gray, Mono, Mono (Rev))
Specifies the color for the created image file.
Image
size
(UnCompressed, Compressed)
Determines whether the image file is compressed.
Save GUI
(Yes, No)
Determines whether the GUI section is also saved.
Wave
Image
(Print
Image)
Output File
(All, 1 to 250)
Specifies the number of files to save.
To specify a range, open the
[Printer]
sheet from the
System screen and set the printing area to
[A-B Wave]
.
(Memory/Recorder Function only)
Wave Image
Shot
(1 to 60 div)
Specifies the data volume per file.
If the number of measurement data is lower than the
number of selected divisions, only the number of
measurement data will be saved. (Memory/Recorder
Function only)
Calc
Result
Folder to
save
(New File, Existing File)
Determines whether data are saved under a new file
name every time (with automatically incremented
number if the same file name exists), or appended to
the same file.
All Blks
Blocks with waveforms are all saved.
Start-End
All used blocks from the start block are saved.
If
[Image size]
is used when saving a display image file, the image may not be
viewable with some image viewer applications.
Summary of Contents for 8847
Page 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Page 2: ......
Page 18: ...Operating Precautions 8 ...
Page 26: ...1 4 Basic Key Operations 16 ...
Page 74: ...4 4 Waveform Observation 64 ...
Page 98: ...5 5 Managing Files 88 ...
Page 110: ...6 6 Miscellaneous Printing Functions 100 ...
Page 126: ...7 8 Seeing Block Waveforms 116 ...
Page 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Page 198: ...10 6 Numerical Calculation Type and Description 188 ...
Page 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Page 304: ...16 2 External I O 294 ...
Page 362: ...Appendix 5 FFT Definitions A28 ...
Page 368: ...Index Index 6 ...
Page 369: ......
Page 370: ......
Page 371: ......