8.5 Converting Input Values (Scaling Function)
131
8
Chapt
er 8
Ut
ilit
y Funct
ions
However, you may need to switch the vertical axis (voltage axis) range to suit
actual input values.
For example, to display ±0.2 V at full scale, set the vertical display to 20 mV per
division (the instrument's 20 mV/div range)
For the rated capacity and rated output, consult the calibration record of the sen-
sor to be used. Set as follows:
With scaling, signals from the sensor are
acquired as current values.
A/B cursors and gauges are displayed
and printed with current (Ampere) val-
ues.
See:
Gauge: (
Cursor A/B value:
Before Scaling
After Scaling
[A]
Using Model 8969 Strain Unit
Example 2
Using the 20 G rated capacity and a sensor with 1000
μ
V/V rated output,
display measured data in units of [G]
Setting Items
Setting Choice
Disp
NUM
Unit
G
Rated Cap
(Rated capacity)
20 [G]
Rated Output
1k
By using the Scaling function, signals
from the sensor are acquired as physi-
cal values.
A/B cursors and gauges are displayed
and printed as physical (G) values.
See:
Gauge: (
Cursor A/B value: (
Before Scaling
After Scaling
[G]
Set the rated capacity (2 x rated output) to be less than ±9.9999±9.
Summary of Contents for 8847
Page 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Page 2: ......
Page 18: ...Operating Precautions 8 ...
Page 26: ...1 4 Basic Key Operations 16 ...
Page 74: ...4 4 Waveform Observation 64 ...
Page 98: ...5 5 Managing Files 88 ...
Page 110: ...6 6 Miscellaneous Printing Functions 100 ...
Page 126: ...7 8 Seeing Block Waveforms 116 ...
Page 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Page 198: ...10 6 Numerical Calculation Type and Description 188 ...
Page 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Page 304: ...16 2 External I O 294 ...
Page 362: ...Appendix 5 FFT Definitions A28 ...
Page 368: ...Index Index 6 ...
Page 369: ......
Page 370: ......
Page 371: ......