5.1 Data capable of Being Saved & Loaded
67
5
Chap
ter
5
Sav
ing
/Loadi
ng
Dat
a &
Managi
ng Fi
les
The following kinds of data can be saved by and loaded into the instrument.
Data Not Loadable on the Instrument ______________________________
• Data saved on devices other than the 8847 Memory HiCorder.
• Image file (
)
•
file
5.1
Data capable of Being Saved & Loaded
Data the Instrument Can Save & Load
O: Possible/ -: Not Possible
File Type
File
Format
Indica-
tion
File Extension & Description
Save
Load
PC
Readable
Auto
Man-
ual
Settings Data
*1
Binary
SET
Settings data
(Measurement Configuration)
-
O
O
-
Waveform Data
*2
Whole of the waveform ac-
quired by the instrument or a
section of the waveform spec-
ified with the A and B
cursors.
Binary
MEM
Memory Function
waveform data
O
O
O
-
*4
REC
Recorder Function
waveform data
O
O
O
-
*4
XYC
X-Y Recorder Function
waveform data
-
O
O
-
FFT
FFT Function data
O
O
O
-
Text
CSV
Text Data
O
O
-
O
Waveform
Management Data
*3
(Divided Saving)
(Index
file)
IDX
Index data for divided saving
O
O
O
-
SEQ
Index data for memory divi-
sion (automatically created
during batch saving)
O
O
O
-
Captured Screen Image
(Display/Waveform
screens)
BMP
*5
BMP Image Data
-
O
-
O
Numerical Calculation
Results
Text
CSV
Text Data
O
O
-
O
Comment for Printing
Text
TXT
Text Data
-
-
O
*6
O
*1: Multiple data can be saved in the instrument, and loaded selectively. Automatic loading at power-on is also possible.
*2: When the data is to be reloaded on the instrument, save it in binary format. Waveforms and some measurement set-
tings are saved.
When the data is to be loaded on a PC, save it in text format. (
When saving a section of a waveform, use the A and B cursors to set the section. (
*3: When you want to use memory division and load all blocks at the same time:
Save measurement data using
[All blocks]
. A directory is automatically created and waveform data for each block
and index data (SEQ) is created. When loading, load this index data.
When loading waveform data of divided saving: Import the IDX index data.
*4: Loading is possible with the Wave Viewer (Wv).
*5: This is a standard Windows graphics format. File in this format can be handled by many graphics programs.
*6: It is possible to print out the text file created by a PC together with loaded waveforms. Other processing operations
are not available.
: File type cannot be handled by 8847.
Files larger than 2 GB cannot be saved.
Summary of Contents for 8847
Page 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Page 2: ......
Page 18: ...Operating Precautions 8 ...
Page 26: ...1 4 Basic Key Operations 16 ...
Page 74: ...4 4 Waveform Observation 64 ...
Page 98: ...5 5 Managing Files 88 ...
Page 110: ...6 6 Miscellaneous Printing Functions 100 ...
Page 126: ...7 8 Seeing Block Waveforms 116 ...
Page 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Page 198: ...10 6 Numerical Calculation Type and Description 188 ...
Page 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Page 304: ...16 2 External I O 294 ...
Page 362: ...Appendix 5 FFT Definitions A28 ...
Page 368: ...Index Index 6 ...
Page 369: ......
Page 370: ......
Page 371: ......