10.2 Settings for Numerical Value Calculation
176
10.2 Settings for Numerical Value Calculation
1
2
Procedure
To open the screen: Press the
STATUS
key
→
[Num Calc]
sheet
1
Enable the Numerical Calculation function.
Move the cursor to the
[Numerical Calc]
item.
Select
[On]
.
2
Specify the Numerical Calculation range.
Move the cursor to the
[Calc Area]
item.
Select
See:
"7.2 Specifying a Waveform Range (A/B Cursor)" (
3
Perform calculation settings.
(1)
Select the calculation type.
Move the cursor to the number of the calculation type for
which to make settings, and select the calculation type.
(When you select
[List]
a list of calculation types appears. To clear the dis-
play, press the
TRIG.SET
key once more.)
WholeWave
Applies calculations to the whole
waveform. (default setting)
A-B Wave
Applies calculations to the data be-
tween A/B cursors.
.Calculation No.
Calculation
Type
(1)
(2)
(3)
Parameters*
*: Setting choices depend
on the calculation type.
Channel for
Calculation
(4)
Statistical
calculation
(5)
Phenomenon
*: Setting can also be made for logic channels
X-Y Area
Area of X-Y composite waveform
Time to Lev*
Time from trigger to specified level
Lev-time
Measurement value at a specified time
point after triggering
Pulse Width*
Pulse width of waveform data
Duty Ratio*
Duty of waveform data
Pulses*
Pulse count of waveform data
Calculation
Four arithmetic operations on numerical
calculation results
Time Diff*
Time difference between phenomenon A
and B.
Phase Diff*
Time difference between phenomenon A
and B displayed as a phase contrast.
High Level
High level value for waveform data
Low Level
Low level value for waveform data
Off
No calculation. (default setting)
Average
Average value of waveform data
Rms
RMS value of waveform data
Peak-Peak
Peak-to-peak value of waveform data
Maximum
Maximum value of waveform data
Max-Time
Time from trigger to maximum value
Minimum
Minimum value of waveform data
Min-Time
Time from trigger to minimum value
Period
Period of signal waveform
Frequency
Frequency of signal waveform
RiseTime
Rise time of waveform data
FallTime
Fall time of waveform data
Std Dev
Standard deviation of waveform data
Area
Area enclosed by zero position and
signal waveform
When selecting
[A-B Wave]
, specify the
calculation range using the A/B cursors on
the Waveform screen.
If no measurement data has been ac-
quired by the instrument, first measure
once so that the range can be specified for
calculations to be applied to subsequent
measurements.
Also select
[Judge]
if you require judg-
ment of calculation results.(
Summary of Contents for 8847
Page 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Page 2: ......
Page 18: ...Operating Precautions 8 ...
Page 26: ...1 4 Basic Key Operations 16 ...
Page 74: ...4 4 Waveform Observation 64 ...
Page 98: ...5 5 Managing Files 88 ...
Page 110: ...6 6 Miscellaneous Printing Functions 100 ...
Page 126: ...7 8 Seeing Block Waveforms 116 ...
Page 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Page 198: ...10 6 Numerical Calculation Type and Description 188 ...
Page 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Page 304: ...16 2 External I O 294 ...
Page 362: ...Appendix 5 FFT Definitions A28 ...
Page 368: ...Index Index 6 ...
Page 369: ......
Page 370: ......
Page 371: ......