17.6 Input Modules Specifications
308
17.6.5 8970 Freq Unit
Temperature and humidity range
for guaranteed accuracy
23 ± 5°C (73 ± 9F°), 20 to 80% RH
Measurement functions
Based on voltage input, measures frequency, rotation speed, power frequency, integral values,
pulse duty ratio, and pulse width.
Connection terminal
Insulated BNC terminal
Input resistance
1 M
Ω
±1%
Input capacitance
30pF±10pF
Maximum input voltage
400 VDC
Maximum rated voltage to earth
300 VAC, DC (Measurement category II), anticipated transient overvoltage 2500 V (between each
input channel and main unit, and between input channels)
Isolation resistance
Dielectric strength
DC2.704 kVAC for 15 seconds (between input module and main unit, and between input modules),
100 M
Ω
and over/500 V
Input type
Unbalanced (input isolated from output)
Guaranteed accuracy period
1 year
Frequency mode
Measurement ranges
1, 5, 10, 50, 100, 500, 1k, 5 kHz/div (f.s.=20div)
Measurement accuracy
±0.1%f.s. (except for 5 kHz/div range) ±0.7%f.s.(5 kHz/div range)
Measurement ranges
DC to 100 kHz (min. pulse width 2
μ
s)
Rotation speed mode
Measurement ranges
100, 500, 1k, 5k, 10k, 50k, 100 kr/min /div (f.s.=20div)
Measurement accuracy
±0.1%f.s. (except for 100 kr/min range) ±0.7%f.s.(100 kr/min range)
Measurement ranges
0 to 2 kr/min (min. pulse width 2
μ
s)
Power frequency mode
Measurement ranges
50 Hz (40 to 60 Hz), 60 Hz (50 to 70 Hz), 400 Hz (390 to 410 Hz) (f.s.=20div)
Measurement accuracy
±0.03 Hz (50 Hz, 60 Hz) ±0.1 Hz (400 Hz)
Integral values mode
Measurement ranges
2k, 10k ,20k ,100k ,200k ,1M counts/div
Measurement accuracy
±range/2000
Measurement ranges
DC to 100 kHz (min. pulse width 2
μ
s)
Pulse duty ratio mode
Measurement ranges
5%/div (f.s.=20div)
Measurement accuracy
±1% (10 to 10 kHz) ±4% (10k to 100 kHz)
Measurement ranges
10 to 100 kHz (min. pulse width 2
μ
s)
Pulse width mode
Measurement ranges
500
μ
, 1m, 5m, 10m, 50m, 100 ms/div (f.s.=20div)
Measurement accuracy
±0.1%f.s.
Measurement ranges
2
μ
to 2 s
Measurement resolution
Integral values mode: 2000LSB/DIV (f.s.=20 DIV)
Power frequency mode: 100LSB/DIV (f.s.=20 DIV)
Except for Integral values mode and Power frequency mode: 500LSB/DIV (f.s.=20 DIV)
Response time
Less than 40
μ
s + sampling interval of instrument in which the device is installed.
Input voltage range
±10 V, ±20 V, ±50 V, ±100 V, ±200 V, ±400 V
Threshold value
±10 V range : -10 to +10 V variable (0.1 V steps), ±20 V range: -20 to +20 V variable (0.2 V steps),
±50 V range : -50 to +50 V variable (0.5 V steps), ±100 V range: -100 to +100 V variable (1 V steps),
±200 V range: -200 to +200 V variable (2 V steps), ±400 V range: -400 to +400 V variable (5 V
steps)
Slope
Rising, falling (Frequency, Rotation speed, Power frequency, Integral values mode)
Level
High, Low (Duty ratio, Pulse width mode)
Hold
Frequency mode/ Rotation speed mode: ON/OFF (1 Hz, 0.5 Hz, 0.2 Hz, 0.1 Hz)
:When OFF is selected and the next measurement value is not determined within the waiting time
period, the frequency or the rotation speed will be determined using the value which is calculated
based on the time period between the timing of the last measurement and the timing of the sampling.
Put it to 0 when the measurement value is below the fixed value.
Smoothing
OFF, ON* (Frequency mode,Rotation speed mode)
*:The permissible smoothing frequency is up to 10 kHz.
Low-pass filter
OFF, 5, 50, 500, 5k, 50 kHz
Input Coupling
DC, AC (Low frequency cut-off in AC-coupled mode : 7 Hz)
Frequency dividing function
Setting range: 1 to 4096, by one step (Frequency ,Rotation speed ,Integral values mode)
Integration start timing
Start, Trigger (Integral values mode)
Integration limit
Hold, Loop (Integral values mode)
Operating temperature and hu-
midity ranges
Same as the Memory HiCorder in which the 8970 is installed
Summary of Contents for 8847
Page 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Page 2: ......
Page 18: ...Operating Precautions 8 ...
Page 26: ...1 4 Basic Key Operations 16 ...
Page 74: ...4 4 Waveform Observation 64 ...
Page 98: ...5 5 Managing Files 88 ...
Page 110: ...6 6 Miscellaneous Printing Functions 100 ...
Page 126: ...7 8 Seeing Block Waveforms 116 ...
Page 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Page 198: ...10 6 Numerical Calculation Type and Description 188 ...
Page 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Page 304: ...16 2 External I O 294 ...
Page 362: ...Appendix 5 FFT Definitions A28 ...
Page 368: ...Index Index 6 ...
Page 369: ......
Page 370: ......
Page 371: ......