9.11 Using trigger settings to search measurement data
171
9
Chapt
er
9
T
rigger
Set
tings
Trigger settings can be used to search measurement data.
Locations that match the set trigger criteria in the measurement data are searched for and displayed
sequentially.
Using the same method to set triggers, it is possible to search for criteria different to those at the
time of measurement.
Example 1: The measurement trigger criteria is trigger level 0 V and the data search condition is
level 100 V.
Example 2: After measurement with a free run without a trigger, locations that exceed 100 V are
searched for.
9.11 Using trigger settings to search measure-
ment data
Procedure
To open the screen: Press the
DISP
key
→
Waveform screen
→
Press the
TRIG.SET
key
→
Trigger settings window
(
[Analog Trig]
, or
[Logic Trig]
sheet)
1
After measurement is finished, set the search criteria.
The setting method is the same as described in "9.3 Trigger-
ing by Analog Signals" (
Logic Signals (Logic Trigger)" (
2
Specify the search range.
Move the cursor to the
[
Area
]
item.
Select
(When using memory division)
Select
3
Execute search.
Move the cursor to the
[
Trig search
]
item.
Select
Whole Wave
Search all measured data. (default setting)
A-B Wave
Search in the range between A/B cursor.
Whole Display
Blocks
Applies search to the whole display blocks.
A/B Display
Blocks
When setting the A/B cursors on the display blocks, ap-
plies search to the range between the A and B cursors on
the display blocks.
Whole Blocks
Applies search to the data in the whole blocks.
A/B Whole
Blocks
When setting the A/B cursors for the whole blocks, applies
search to the range between the A and B cursors for the
whole blocks.
Execute
search
Start the search from the beginning of the measurement
data.
Search next
Search for the next search condition from the current po-
sition.
Move A cursor Move A cursor to the searched position.
Move B cursor Move B cursor to the searched position.
Clear
Clear the searched position.
3
2
Search and search for next can be per-
formed using the
T
(manual trigger)
key as well.
•
When selecting A/B range and only
A cursor is shown on the screen,
search will be performed on the
data after A cursor.
•
The search range for
[Whole
Blocks]
and
[A/B Whole Blocks]
will be defined according to the
start block and the number of
blocks to be measured.
Summary of Contents for 8847
Page 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Page 2: ......
Page 18: ...Operating Precautions 8 ...
Page 26: ...1 4 Basic Key Operations 16 ...
Page 74: ...4 4 Waveform Observation 64 ...
Page 98: ...5 5 Managing Files 88 ...
Page 110: ...6 6 Miscellaneous Printing Functions 100 ...
Page 126: ...7 8 Seeing Block Waveforms 116 ...
Page 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Page 198: ...10 6 Numerical Calculation Type and Description 188 ...
Page 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Page 304: ...16 2 External I O 294 ...
Page 362: ...Appendix 5 FFT Definitions A28 ...
Page 368: ...Index Index 6 ...
Page 369: ......
Page 370: ......
Page 371: ......