9.3 Triggering by Analog Signals
156
3. Voltage Sag Trigger (
only) _____________________________
Triggering occurs when the voltage peak drops below a preset level for more
than half a cycle. The allowable time axis range is 20
μ
s - 50 ms/div.
Procedure
To open the screen: Press the
DISP
key
→
Waveform screen
→
Press the
TRIG.SET
key
→
Trigger settings
window (
[Analog Trg.]
sheet)
1
Move the cursor to the
[Type]
item of the channel for which to make the setting.
2
Select the trigger type with the
F
keys.
3
Move the cursor to the parameter item with the
CURSOR
keys.
4
Select the parameter value with the
F
keys.
1
3
1. Level Trigger (
2. In-Window Trigger
Out-of-Window Trigger (
3. Voltage Sag Trigger (
4. In-Period Trigger
Out-of-Period Trigger (
5. Glitch Trigger (
Type
Parameters
[VDrop]
[L]
(Level)
Sets the level (voltage value) for the trigger. (The setting
can be made in 1/50 increments.)
[f=]
(Frequency) Select 50 or 60 Hz.
T
1/2 Period
Trigger Level
Summary of Contents for 8847
Page 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Page 2: ......
Page 18: ...Operating Precautions 8 ...
Page 26: ...1 4 Basic Key Operations 16 ...
Page 74: ...4 4 Waveform Observation 64 ...
Page 98: ...5 5 Managing Files 88 ...
Page 110: ...6 6 Miscellaneous Printing Functions 100 ...
Page 126: ...7 8 Seeing Block Waveforms 116 ...
Page 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Page 198: ...10 6 Numerical Calculation Type and Description 188 ...
Page 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Page 304: ...16 2 External I O 294 ...
Page 362: ...Appendix 5 FFT Definitions A28 ...
Page 368: ...Index Index 6 ...
Page 369: ......
Page 370: ......
Page 371: ......