8.10 Making Detailed Settings for Input Modules
144
The 8969 Strain Unit can perform auto balance.
When auto balance is performed, the reference output level of the conversion
unit can be matched with the specified zero position.
It is applicable only to a 8969 Strain Unit.
Before executing auto-balance ___________________________________
• Turn power on and wait 30 minutes to allow the internal temperature of the
input module to stabilize.
• With the sensor connected to the input module, execute auto-balance under
stable input conditions.
• Auto-balance cannot execute during measurement.
• Key operations are not accepted while auto-balance is executing.
To execute auto-balance ________________________________________
See:
Opening the
[Each Ch]
sheet, Making a Channel Selection (
Auto Balance can also be executed from the Channel settings window (Analog
sheet) (if the range of a channel with installed 8969 Strain Unit is selected).
See:
Opening the Channel settings window (
[Analog]
sheet): (
In the following cases, auto-balance should be executed again.
• After changing the vertical axis (strain axis) range
• After an input module has been removed or inserted
• After the strain gauge transducer has been replaced
• After power has been turned off and on
• After performing a system reset
• When ambient temperature has changed significantly (the zero position may
drift)
8.10.4 Settings for the 8969 Strain Unit
Setting Item:
[Range]
Selections
Description
Auto Bal All
Chs
Auto Balance is executed for all channels where the 8969 Strain
Unit is installed.
Auto Bal Ch 1
Auto Balance is executed for the currently selected channel.
If "Warning: Auto balance failed." appears:
The channel on which auto-balance failed is displayed.
Verify the following, and execute again:
• Is the sensor in a discharged state? (Make sure that it is not being subject to
vibration, etc.)
• Is the sensor connected correctly?
Summary of Contents for 8847
Page 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Page 2: ......
Page 18: ...Operating Precautions 8 ...
Page 26: ...1 4 Basic Key Operations 16 ...
Page 74: ...4 4 Waveform Observation 64 ...
Page 98: ...5 5 Managing Files 88 ...
Page 110: ...6 6 Miscellaneous Printing Functions 100 ...
Page 126: ...7 8 Seeing Block Waveforms 116 ...
Page 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Page 198: ...10 6 Numerical Calculation Type and Description 188 ...
Page 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Page 304: ...16 2 External I O 294 ...
Page 362: ...Appendix 5 FFT Definitions A28 ...
Page 368: ...Index Index 6 ...
Page 369: ......
Page 370: ......
Page 371: ......