12.2 Display Settings
205
12
Chapte
r 12
Mem
o
ry Div
is
ion Funct
ion
Getting Details on Each Block:
The trigger time and measurement status of each block can be viewed on the list.
Move the cursor to the
[Map/List]
, and select
[List]
.
Block No.
A block can be selected by the
CURSOR
keys or
F
keys.
You can move the cursor to the Reference Block column to set a block’s on/off state as a Refer-
ence Block.
To switch block waveforms on the Waveform screen:
To be able to select the block you want to see, press the
WAVE
key at the Waveform screen and
switch Pos to Block.
See:
"7.8 Seeing Block Waveforms" (
• When displaying memory division blocks as a list, blocks may have the same
trigger times. This occurs because the minimum resolution of the clock used
by this unit is 1/128th of a second (7.8125 ms) and measurement occurs dur-
ing this interval.
• If triggers occur continuously within an interval shorter than 500
μ
s, the
displayed trigger time may indicate a time slower than reality.
Example:
The recording time is 1.25 ms and the trigger
times of No. 3 and No. 4 blocks may be the
same.
Summary of Contents for 8847
Page 1: ...MEMORY HiCORDER Instruction Manual 8847 December 2010 Revised edition 4 8847A981 04 10 12H ...
Page 2: ......
Page 18: ...Operating Precautions 8 ...
Page 26: ...1 4 Basic Key Operations 16 ...
Page 74: ...4 4 Waveform Observation 64 ...
Page 98: ...5 5 Managing Files 88 ...
Page 110: ...6 6 Miscellaneous Printing Functions 100 ...
Page 126: ...7 8 Seeing Block Waveforms 116 ...
Page 160: ...8 10 Making Detailed Settings for Input Modules 150 ...
Page 198: ...10 6 Numerical Calculation Type and Description 188 ...
Page 296: ...15 7 Controlling the Instrument with Command Communications LAN USB 286 ...
Page 304: ...16 2 External I O 294 ...
Page 362: ...Appendix 5 FFT Definitions A28 ...
Page 368: ...Index Index 6 ...
Page 369: ......
Page 370: ......
Page 371: ......