Event Outputs
Output channels
4 digital isolated output channels (open Collector, Emitter)
Only supported by Digital Event/Timer/Counter connector
Isolation device
Vishay VOS617A optocoupler (or comparable)
Output frequency
170 kHz output signal tested.
Maximum useable frequency for the system is limited by the Isolated Digital Event Adapter
or acquisition system, whichever is the slowest.
Nondestructive control voltages (see Figure B.64 and Figure B.65)
Maximum voltage
0.007 * R
ext
and < 80 V
Minimum voltage
-
7.0 V
Isolated Output
(only from mainframe digital IO)
C
+
–
U
control
Data
E
I
f
430Ω
VOS617A
I
c
Mainframe d
igital IO
Rext
Figure B.64:
Add an external resistor to change output voltage levels
10
0
-10
20
30
40
50
60
70
80
90
0
4
6
2
8
10
12
14
U
external
[V]
R
ext
[kΩ]
Maximum
Minimum
Figure B.65:
Nondestructive output voltages
Output Connectors (Only one output connector option can be used at the same time)
Digital Event/Timer/Counter
HD22 sub
-
D 44 pin male (connecting cable included)
Digital marker group A and B
15 pin, female sub
-
D type (connecting cable 1
-
KAB2116
-
1.5, to be ordered separately)
Output Cables
GEN DAQ mainframe to adapter
0.7 m (2.30 ft), included with adapter
GEN DAQ marker card to adapter
1
-
KAB2116
-
1.5, 1.5 m (4.92 ft) needs to be ordered separately
Power
GEN DAQ mainframe to adapter
Powered by GEN DAQ mainframe or marker card
GEN3i
778
I3763-3.1 en HBM: public