14
In the drop
-
down list that comes up, select
Transceive
for all recorders.
Figure E.87:
Master/Slave trigger list
15
Set the filter mode of both channels to "Wideband".
16
Set up a display with the first channel of the first recorder card in the master
mainframe and the first channel of the first recorder card in the slave
mainframe.
Making a multi
-
mainframe recording
1
Wait for the “MASTER SLAVE” status to display
Synchronized
before
proceeding to the next step.
Figure E.88:
MASTER SLAVE Synchronized
2
Press
Run
in the acquisition control panel to start a recording.
3
The signal on the master mainframe will now generate a trigger event. This
trigger event will be relayed to the slave mainframe.
4
The recording will now show the rising edge of the TTL level 1 Hertz signal
recorded by the master mainframe and the slave mainframe.
5
The recordings in both mainframes are started at the same time.
6
All recorded signals will match in time to within
±
150 ns.
If all signals match in time, the recordings were completed successfully.
GEN3i
I3763-3.1 en HBM: public
903