B
Timer Counter mode:
Count Quadrature
-
> counters monitor the
transition of the four different states the signal can be in.
C
Reset Mode:
Each External Pulse
-
> resets the counter every time we
receive a pulse in Reset input.
D
Range from/to
0 to 360°
E
Technical units multiplier:
87.891
m°/Count
This setting assumes the use of an encoder with
4096
pulses/rotation
1024 pulses [from encoder] x 4 transitions [quad precision selected].
F
Technical units offset:
0°
G
Technical units:
° (degrees)
H
Invert reset pin:
Deactivated*
-
> A High Level of Reset input is
needed to reset the counter to 0.
Note
* The counter works ONLY when the reset input is set to Low. After you reset
the counter with a High Level, you need to return the reset level to Low to
continue the measurement. In other words, your reset signal needs to be a
really short impulse in order to minimize the amount of time that the counter is
not counting.
GEN3i
I3763-3.1 en HBM: public
305