10.4.5
Event counter
Sometimes it is not possible to trigger on a specified condition using a selected
trigger mode alone, because several events meet the required situation. So far,
we have seen “filters” that can be used to narrow the range of trigger
candidates, like hold off and interval timer.
As a last resource, the event counter can be used. The event counter counts
all channel triggers that have been detected and generates a system trigger
when the count equals a user defined value.
GEN3i
I3763-3.1 en HBM: public
161