G.3.3
GEN3i/GEN7i with up to three QuantumX "B" version modules
Figure G.10:
GEN3i with a mix of three QuantumX "B" version modules
Note
Setup only shows GEN3i. GEN3i can be replaced by GEN7i.
Note
* All modules must be “B“ versions; one module must set to “clock master“ and
connected to the GEN3i PTP Ethernet.
HINT/TIP
This setup uses multiple network ports on the GEN3i PC section.
Use fixed IP address setup with different base IP address and non overlapping
net masks for each of the GEN3i network ports to make sure the setup always
works.
Background network details
If the network ports are configured for DHCP setup, each of the connections
using the APIPA protocol to find a free IP address. As none of the network ports
are linked together the APIPA protocol does not detect the address used any
of the network devices in this setup. At random all QuantumX systems might
end up with exactly the same network IP address. If this happens the systems
are not uniquely addressable anymore and the communication fails.
GEN3i
I3763-3.1 en HBM: public
915