l
Select
Redo Verification
to restart the jitter detection.
l
Select
Exit Program
to stop the touch screen validation. When exit is
selected, no calibration adjustment has been performed.
l
Select
Next Step
to continue to the calibration steps. Do not progress
to the next step if this step fails.
2
Twenty (20) points need to be tapped on a grid. The tap positions are
indicated on the screen by a crosshair.
The crosshair starts at the top left point, moves from left to right and then
to the next row of points below until all points have been tapped. A margin
is applied to detect whether the tap is in the immediate area around the
point that is expected to be tapped.
Figure E.33:
Touch screen calibration
-
Grid with crosshair (step 3)
A Crosshair
Moved from start point in the top left
-
hand corner.
B Description
To provide calibration samples, tap the crosshair each
time that it appears on the screen.
Right
-
click anywhere on the screen to return to the last calibration
point. Press the Esc key to close the tool. Do not change the screen
orientation until the calibration process is completed.
3
After all points on the grid have been tapped, the calibration data can be
saved. If the calibration was successful, select
Save
to save the new data.
GEN3i
844
I3763-3.1 en HBM: public