When in Single sweep, Multiple sweep or Slow
-
Fast Sweep mode, the start and
stop synchronization of the recording is irrelevant. The entire recorded sweep
data is determined by the trigger origin with a fixed pre
-
and post
-
trigger time
frame. In sweep
-
based recordings, the acquisition system is typically started
first. Checks on all system parts are then performed and the first trigger is
inserted only when all systems are ready.
Mainframe 1
Mainframe 2
Mainframe N
t
start
t
phase
t
stop
Figure 14.8:
Measuring start/stop synchronization accuracy
GEN3i
I3763-3.1 en HBM: public