11.3.4
Setting the Master/Slave trigger
When the Master/Slave card is in use, a recorder can either put the recorder
trigger on the Master/Slave trigger line and/or pick up the trigger from the
Master/Slave trigger line.
There are four settings that can be selected in the Perception software:
l
Disabled
No trigger is transmitted to or received from other mainframes.
l
Transmit
Transmit trigger(s) from this recorder to other mainframes, received triggers
are ignored.
l
Receive
Receive trigger(s) from other mainframes, no triggers are transmitted.
l
Transceive
Transmit and receive trigger(s) from other mainframes.
The settings are controlled in the block diagram or through the
Master/Slave
trigger
setting in the sheet.