Analog Input Section
Common mode (referred to system ground)
Ranges
Less than
±
2 V
Larger than or equal to
±
2 V
Rejection (CMR)
> 80 dB @ 80 Hz (100 dB typical)
> 60 dB @ 80 Hz (80 dB typical)
Maximum common mode voltage
2 V RMS
33 V RMS
-120
-100
-80
-60
-40
-20
0
0.01
0.1
1
10
100
1000
Frequency [kHz]
Common mode response
< ± 2 V ranges
≥ ± 2 V ranges
M
ag
n
it
u
d
e
[d
B
]
Figure A.278:
Representative common mode response
Input overload protection
Over voltage impedance change
The activation of the over voltage protection system will result in a reduced input impedance.
The over voltage protection will not be active as long as the input voltage is less than
200% of the selected input range or 50 V DC whichever is the smallest value.
Maximum nondestructive voltage
±
50 V DC
Overload recovery time
Restored to 0.1% accuracy in less than 5 μs after 200% overload
Input Ranges When Using Passive Voltage Probes
Detailed probe specifications can be found at the end of this datasheet
Single
-
ended
Added voltage ranges
G901 (10:1 divide factor)
±
50 V,
±
100 V,
±
200 V
G902 (10:1 divide factor)
±
50 V,
±
100 V,
±
200 V
G903 (100:1 divide factor)
±
50 V,
±
100 V,
±
200 V,
±
500 V,
±
1 kV
G904 (100:1 divide factor)
±
50 V,
±
100 V,
±
200 V,
±
500 V,
±
1 kV,
±
2 kV
G906 (1000:1 divide factor)
±
50 V,
±
100 V,
±
200 V,
±
500 V,
±
1 kV,
±
2 kV,
±
5 kV,
±
10 kV (
±
20 kV @ DC to
60 Hz)
Differential matched
Added voltage ranges
G907 (10:1 divide factor)
±
50 V,
±
100 V,
±
200 V
Input Ranges When Using Active Differential Voltage Probes
G909 (20:1 divide factor)
±
140 V RMS input and
±
1000 V RMS common mode
G909 (200:1 divide factor)
±
1000 V RMS input and
±
1000 V RMS common mode
Input Ranges When Using Current Clamps
Detailed probe specifications can be found at the end of this datasheet
Clamp type
Added current ranges
G912 (AC/DC)
±
30 mA to
±
30 A DC
±
30 mA to
±
20 A RMS
G913 (AC)
±
100 mA to
±
1000 A RMS
G914 (AC)
±
50 mA to
±
20 A RMS
GEN3i
I3763-3.1 en HBM: public