Wideband (No Anti
-
Alias Protection)
When wideband is selected, there is neither an analog anti
-
alias filter nor any digital filter in the signal path. Therefore, there is no anti
-
alias
protection when wideband is selected.
Wideband bandwidth
Between 540 kHz and 690 kHz (
-
3 dB)
Passband flatness
(1)
0.1 dB; DC to 150 kHz
Wideband overview
10
0
-10
-20
-30
-40
-50
-60
-70
-80
Frequency [kHz]
Ma
gni
tu
de
[dB
]
10000
1000
100
10
1
0.1
1000
100
10
1
0.1
Passband flatness
0.2
0.1
0
0.05
0.15
-0.1
-0.15
-0.05
-0.2
Frequency [kHz]
Ma
gni
tu
de
[dB
]
Figure L.11:
Representative Wideband examples
(1)
Measured using a Fluke 5700 calibrator, DC normalized
GEN3i
I3763-3.1 en HBM: public
1039