WARNING
High voltage and qualified personnel
For measurements falling within the scope of the EN 50110
-
1 and
EN 50110
-
2, please note that all cards with working voltages above
50 V AC RMS or 120 V DC may only be connected by a qualified technician or
a person trained in electrical engineering and supervised by a qualified
technician. (Qualified technicians are persons who, due to their specialist
training, knowledge and experience, as well as their knowledge of the relevant
provisions, are able to assess the work with which they are entrusted and detect
possible risks and who have been nominated as qualified technicians by their
employer).
WARNING
Connectors and cables
Do not use non
-
protected or non
-
shrouded connectors.
The following connectors are not safe to use with this card and must not be
used (see Figure 12.4).
Figure 12.4:
Unsafe connectors
The inputs on the 1kV card are compatible with the following connectors and
cables (see Figure 12.5). All cables used with the 1kV card must support
1000 V DC (or 1000 V AC peak) and 600 V CAT II. All required cables and
connectors can be found in the appendix “Options, to be ordered separately”
on page 728.
GEN3i
218
I3763-3.1 en HBM: public