Acquisition Modes
Single sweep
Triggered acquisition to on
-
board memory without sample rate limitations; for single
transients or intermittent phenomena. No aggregate sample rate limitations.
Multiple sweeps
Triggered acquisition to on
-
board memory without sample rate limitations; for repetitive
transients or intermittent phenomena. No aggregate sample rate limitations.
Slow
-
Fast Sweep
Identical to single sweep acquisition with additional support for fast sample rate switches
during the post
-
trigger segment of the slow rate single sweep settings. No aggregate sample
rate limitations. Slow
-
Fast Sweep is not supported by the RT
-
FDB calculators.
Continuous
Direct storage to PC or mainframe controlled hard disk without file size limitations; triggered
or un
-
triggered; for long duration recorder type applications. Aggregate sample rate
limitations depend on Ethernet speed, PC used and data storage media used.
Dual
Combination of Multiple sweeps and Continuous; recorder type streaming to hard disk with
simultaneously triggered sweeps in on
-
board memory. Aggregate sample rate limitations
depend on Ethernet speed, PC used and data storage media used.
In Dual mode the RT
-
FDB calculators sample based results are only calculated for the
sweep sections of the recorded data. Due to the asynchronous nature of cycle based results,
all cycle based results are continuously stored and used in both the sweep as well as the
continuous sections of the recording.
Acquisition Mode Details
16 Bit Resolution
Recording Mode
Single Sweep
Multiple Sweeps
Slow
-
Fast Sweep
Continuous
Dual Rate
Enabled channels
Enabled channels
Enabled channels
1 Ch
8 Ch
8 Ch &
events
1 Ch
8 Ch
8 Ch &
events
1 Ch
8 Ch
8 Ch &
events
Max. sweep memory
100 MS
12 MS
10.5 MS
not used
80 MS
9.5 MS
8 MS
Max. sweep sample rate
200 kS/s
not used
200 kS/s
Max. continuous FIFO
not used
100 MS
12 MS
10.5 MS
20 MS
2 MS
2 MS
Max. continuous sample
rate
not used
200 kS/s
Sweep sample rate / 2
Max. aggregate
continuous streaming rate
not used
0.2 MS/s
1.6 MS/s
1.8 MS/s
0.1 MS/s
0.8 MS/s
0.9 MS/s
0.4 MB/s
3.2 MB/s
3.6 MB/s
0.2 MB/s
1.6 MB/s
1.8 MB/s
18 Bit Resolution
Recording Mode
Single Sweep
Multiple Sweeps
Slow
-
Fast Sweep
Continuous
Dual Rate
Enabled channels
Enabled channels
Enabled channels
1 Ch
8 Ch
8 Ch &
events &
Timer/
Counter
1 Ch
8 Ch
8 Ch &
events &
Timer/
Counter
1 Ch
8 Ch
8 Ch &
events &
Timer/
Counter
Max. sweep memory
50 MS
6 MS
4 MS
not used
40 MS
4.5 MS
3 MS
Max. sweep sample rate
200 kS/s
not used
200 kS/s
Max. continuous FIFO
not used
50 MS
6 MS
4 MS
10 MS
1 MS
0.7 MS
Max. continuous sample
rate
not used
200 kS/s
Sweep sample rate / 2
Max. aggregate
continuous streaming rate
not used
0.2 MS/s
1.6 MS/s
2.2 MS/s
0.1 MS/s
0.8 MS/s
1.1 MS/s
0.8 MB/s
6.4 MB/s
8.8 MB/s
0.4 MB/s
3.2 MB/s
4.4 MB/s
Single Sweep
Pre
-
trigger segment
0% to 100% of selected sweep length
If trigger occurs before the pre
-
trigger segment is recorded, the pre
-
trigger segment is
truncated to recorded data only.
Delayed trigger
Maximum 1000 seconds after a trigger occurred. The sweep is recorded immediately after
a delayed trigger time with 100% post
-
trigger after this time point.
GEN3i