4
To enable the recovery mode, use a pointy object (e.g. screwdriver) and
move the lever to the side labeled
Min
.
Figure E.83:
Lever to switch the boot mode
A
Lever
5
Re
-
insert all acquisition cards.
6
Insert the power cord and power on the mainframe.
7
Use Perception software to perform a firmware upgrade.
8
When Perception completes the mainframe will automatically reboot.
9
When reboot is completed turn of the mainframe and remove the power
cord.
10
Remove all acquisition cards to allow access to the mainframes backplane.
11
Shift the black lever to the side labeled
Normal
.
12
Re
-
insert all acquisition cards.
13
Insert the power cord and power on the mainframe.
The GENDAQ mainframe should now work in normal mode.
Note
If the mainframe still fails to work correctly after this recovery process the unit
must be send back to HBM service for a repair.
GEN3i
896
I3763-3.1 en HBM: public