In this situation, a sensor is used with two (out of four) resistors: R4 and R3.
These resistors are placed between A
-
C and C
-
D.
Two additional resistors must be provided: R1 and R2. There are two ways of
doing this:
1
Use the standard 100 kΩ resistors inside the acquisition card.
2
Provide two resistors.
Nothing needs to be done to the hardware for the first situation.
Two resistors need to be added to the plug
-
in module in the locations marked
HB+ and HB
-
for the second situation. You will also need to remove the jumper
J1. Please refer to Figure 12.31 on page 250 for electrical/schematic details
and to Figure 12.32 on page 252 for mechanical/location details.
At least three wires are necessary to connect a bridge such as this. Please refer
to Figure 12.36 for connection details. Inform the Perception software when a
half bridge is used.
To select half
-
bridge completion in Perception
To select half
-
bridge completion in Perception, proceed as follows:
1
In Perception, go to the
Settings
sheet.
2
In the task pane, select the
Bridge
in the
Input
section.
3
Select one or more channels.
GEN3i
260
I3763-3.1 en HBM: public