Wideband (No Anti
-
Alias protection)
When wideband is selected, there is neither an analog anti
-
alias filter nor any digital filter in the signal path. Therefore, there is no anti
-
alias
protection when wideband is selected.
Wideband bandwidth
Between 900 kHz and 1500 kHz (
-
3 dB)
0.1 dB passband flatness
DC to 160 kHz
(1)
100
Magnitude [%]
102.33
101.74
101.16
100.58
99.43
98.86
98.29
97.72
100
102.33
101.74
101.16
100.58
99.43
98.86
98.29
97.72
Magnitude [%]
Frequency [kHz]
Magnitude
[dB
]
1000
100
10
1
0.1
± 10 V: Wideband
passband flatness
0.2
Specification line
Specification line
0.1
0
0.05
0.15
-0.1
-0.15
-0.05
-0.2
Frequency [kHz]
Magnitude
[dB
]
± 10 V: Wideband overview
Frequency [kHz]
Magnitude
[dB
]
1000
100
10
1
0.1
± 2 V: Wideband
passband flatness
0.2
0.1
0
0.05
0.15
-0.1
-0.15
-0.05
-0.2
Frequency [kHz]
Magnitude
[dB
]
± 2 V: Wideband overview
-130
-110
-90
-70
-50
-30
-10
10
0.1
1
10
100
1000
10000
-130
-110
-90
-70
-50
-30
-10
10
0.1
1
10
100
1000
10000
Figure A.120:
Representative Wideband examples
(1)
Measured using a Fluke 5700A calibrator, DC normalized
GEN3i
I3763-3.1 en HBM: public
529