Basic Power Accuracy
The GN610B is calibrated and checked at 53 Hz voltage and current inputs using burden resistors. During calibration burden resistors are
attached to three voltage channels to enable current measurements.
Specifications are given for the 2.5 Ω burden. Using the 1.0 Ω or 10.0 Ω burden will give different current ranges but identical results.
2.5 Ω
Burden
spans
1.264 A DC
800 mA DC
400 mA DC
160 mA DC
80 mA DC
40 mA DC
0
-
100 Hz
Sine wave
CF: 1.41
Cos Phi : 1
Burden
ranges
440 mA RMS
280 mA RMS
140 mA RMS
56 mA RMS
28 mA RMS
14 mA RMS
IT200 ranges
N/A
200 A RMS
(1)
140 A RMS
56 A RMS
28 A RMS
14 A RMS
IT400 ranges
N/A
400 A RMS
(1)
280 A RMS
112 A RMS
56 A RMS
28 A RMS
IT700 ranges
700 A RMS
(1)
490 A RMS
245 A RMS
98 A RMS
49 A RMS
24.5 A RMS
Voltage
spans
Voltage
ranges
Typical
Typical
Typical
Typical
Typical
Typical
40 V DC
14.1 V RMS
0.02% reading
+ 0.05% range
0.02% reading
+ 0.05% range
0.02% reading
+ 0.05% range
0.02% reading
+ 0.1% range
0.02% reading
+ 0.1% range
0.02% reading
+ 0.15% range
100 V DC
35.3 V RMS
0.02% reading
+ 0.05% range
0.02% reading
+ 0.05% range
0.02% reading
+ 0.05% range
0.02% reading
+ 0.1% range
0.02% reading
+ 0.1% range
0.02% reading
+ 0.15% range
200 V DC
70.7 V RMS
0.02% reading
+0.05% range
0.02% reading
+ 0.05% range
0.02% reading
+ 0.05% range
0.02% reading
+ 0.1% range
0.02% reading
+ 0.1% range
0.02% reading
+ 0.15% range
400 V DC
141 V RMS
0.02% reading
+ 0.05% range
0.02% reading
+ 0.05% range
0.02% reading
+ 0.05% range
0.02% reading
+ 0.1% range
0.02% reading
+ 0.1% range
0.02% reading
+ 0.15% range
1 kV DC
353 V RMS
0.02% reading
+ 0.05% range
0.02% reading
+ 0.05% range
0.02% reading
+ 0.05% range
0.02% reading
+ 0.1% range
0.02% reading
+ 0.1% range
0.02% reading
+ 0.15% range
2 kV DC
707 V RMS
0.02% reading
+ 0.05% range
0.02% reading
+ 0.05% range
0.02% reading
+ 0.05% range
0.02% reading
+ 0.1% range
0.02% reading
+ 0.1% range
0.02% reading
+ 0.15% range
(1)
With the right setup the LEM sensor can be used for this range, however this input range is not specified in the LEM current sensor
data sheet.