B.15.2
Passive, single
-
ended isolated probes
G057: Passive, Single
-
Ended Isolated Voltage Probe (Option, to be ordered separately)
To be used with single
-
ended amplifiers or with differential amplifiers in single
-
ended mode
10 * Udiff
Uiso
Uiso
Udiff
(+)
(-)
(-)
GND
GND
Single-ended isolated or
unbalanced differential amplifier
99M
900K
Isolation
Barrier
100K
(+)
(+)
(-)
Uiso
Non-isolated
system
Figure B.93:
Block diagram passive, single
-
ended isolated voltage probe
Isolation
Supported if the acquisition card uses isolated amplifiers
Capacitive compensation range
30 to 70 pF
DC In
-
accuracy
2%
Divide factors
100:1
Probe impedance (connected to channel)
100 MΩ
-
3 dB Bandwidth
50 MHz
Maximum input voltage
600 V RMS CAT III, 1000 V RMS CAT II, 3540 V RMS CAT I
Probe cable length
1.2 m (3.9 ft)
Probe operating temperature range
0 °C to +50 °C (32 °F to 122 °F)
Original manufacturer's part number
Multi
-
Contact Isoprobe II
-
100:1 55pF
Figure B.94:
Probe and probe accessories
GEN3i
796
I3763-3.1 en HBM: public